AssayMAP Protein Sample Prep Workbench : IMAC Cartridge Customization v1.1 User Guide : Assay development guidelines and protocol notes

Assay development guidelines and protocol notes
This topic explains each step of the protocol so that you can optimize the protocol to your particular experimental design.
Note: For guidelines on performing a mock run, see About performing a mock run (optional).
Protocol stepwise guidelines
 
*Cartridges ship dry and, therefore, contain air entrained in the cartridge bed. Failure to prime the cartridges can prevent the reagents and buffers from accessing parts of the resin bed, resulting in reduced capacity and poor reproducibility.
 
Number of Full Columns of Cartridges
This setting is critical to set the proper force used to mount the cartridges. To obtain expected instrument performance, ensure that the column selection is correct. If the column selection is:
Greater than the actual number of columns used, the Bravo Platform will apply too much force when mounting the cartridges, which can cause damage to both the cartridges and the AssayMAP syringes in the head.
For example, if the software specifies 10 columns, but only 1 column of cartridges are in the seating station, the head will apply 12 times more force than what is required.
Less than the actual number of columns used, the Bravo Platform will not apply enough force to seat the cartridges properly.
For example, if the software specifies 1 column, but 12 columns of cartridges are in the seating station, the head will apply 1/12th the force required to seat the cartridges properly. In this case, cartridges may fall off during the run or the volume of liquid that moves across the cartridge bed may be variable do to liquid moving past the syringe cartridges seal into the cartridge cup.
Default: 1
Range: 1-12
Initial Syringe Wash step
The first stage in the protocol is to set up the system and wash the syringes in the Bravo 96AM Head to ensure no carryover from previous experiments remains. The Initial Syringe Wash step removes any potential contaminants from the syringes. Each cycle of the Initial Syringe Wash step aspirates 250 µL into the syringes from the wash station chimneys, and then moves by a fixed offset between the chimneys to dispense the syringe contents to waste.
Select the Initial Syringe Wash step to clean the syringes. Skip this step if you are certain that the syringes are clean. In the default protocol, this step is selected.
Table Setting the Initial Syringe Wash parameter
 
Prime step
The Prime step removes entrained air from within the cartridges and properly wets the surface of the resin. For the most effective priming of AssayMAP Fe(III)-NTA cartridges, the Priming Buffer requires that the solution contain at least 50% organic solvent, typically, acetonitrile. Higher concentrations of organic solvent are also acceptable.
The Prime step aspirates the specified volume of Priming Buffer (deck location 4) into the syringes, mounts the cartridges (deck location 2) onto the syringes, and then dispenses buffer at the specified flow rate through the cartridges into the Waste plate (deck location 3). After removing the cartridges at the seating station (deck location 2), the syringes are washed the specified number of cycles at the wash station (deck location 1) using 250 µL per wash cycle.
Always perform the Prime step at the beginning of a run. In the default protocol, this step is selected.
Table Setting the Prime step parameters
 
Strip Metal step
The Strip Metal step removes the metal coordinated to the Fe(III)-NTA resin by using a strong chelator, such as EDTA. Complete removal of the bound metal from the resin bed is critical so that a different metal can be bound at the maximum metal-binding capacity of the cartridge.
The Strip Metal step aspirates the specified volume of Metal Stripping Buffer (deck location 5) into the syringes, and then mounts the cartridges at the seating station (deck location 2). The buffer is dispensed through cartridges at the specified flow rate into the Waste plate (deck location 3), or into the Flow Through Collection plate (deck location 7) if the Collect Flow Through step is selected. During an external cartridge tip wash, the cartridge tips dip into the wash station chimneys (deck location 1) to remove any residual liquid on the outside of the cartridge. After removing the cartridges at the seating station (deck location 2), the syringes are washed the specified number of cycles at the wash station (deck location 1) using 250 µL per wash cycle.
Always perform the Strip Metal step after cartridge priming to remove bound metal from the Fe(III)-NTA resin. In the default protocol, this step is selected.
Table Setting the Strip Metal step parameters
 
Collect Flow Through (Strip Metal step)
Note: The Collect Flow Through step is an option only if the Strip Metal step is conducted.
If you select the Collect Flow Through step, the flow-through from the Strip Metal step is dispensed directly into the Flow Through Collection plate (deck location 7).
If the Strip Metal step is conducted, but the Collect Flow Through step is not selected, the flow-through is dispensed directly into the Waste plate (deck location 3).
For the maximum practical working volumes per well for specific labware at deck location 7, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
Select the Collect Flow Through step if you want to collect the flow-through generated during the Strip Metal step for downstream processing or for analysis.
Cup Wash 1 step
The Cup Wash 1 step removes the small volume of residual liquid that might remain above the upper cartridge membrane after the Strip Metal step. The Cup Wash step effectively rinses the cup portion of AssayMAP cartridges with buffer. The Cartridge Wash Buffer 1 (deck location 6) is aspirated into the syringes, and then dispensed into the cartridge cups (deck location 2). This liquid, plus any residual liquid, is aspirated from the cartridge cups into the syringes. The protocol ensures that no cartridges are stuck to the probes before dispensing the liquid into the Waste plate (deck location 3), and then washing the syringes at the wash station (deck location 1).
Select the Cup Wash 1 step if residual liquid remains in the cartridge cups from the Strip Metal step. In the default protocol, this step is selected.
Table Setting the Cup Wash 1 step parameters
 
Internal Cartridge Wash 1 step
The Internal Cartridge Wash 1 step aspirates Cartridge Wash Buffer 1 (deck location 6) into the syringes. After mounting the cartridges at the seating station (deck location 2), the buffer is dispensed through the cartridges at the specified flow rate into the Waste plate (deck location 3), or into the Flow Through Collection plate (deck location 7) if the Collect Flow Through step is selected. An external cartridge tip wash is performed at the wash station (deck location 1), where the cartridge tips dip into the chimneys to remove any Cartridge Wash Buffer 1 on the outside of the cartridge. After removing the cartridges at the seating station (deck location 2), the syringes are washed the specified number of cycles at the wash station (deck location 1) using 250 µL per wash cycle.
Select the Internal Cartridge Wash 1 step to flush away any remaining Metal Stripping Buffer and prepare the cartridge with a buffer compatible with the Metal Reagent in the Load Metal step. In the default protocol, this step is selected.
Table Setting the Internal Cartridge Wash 1step parameters
 
Collect Flow Through (Internal Cartridge Wash 1 step)
Note: The Collect Flow Through step is an option only if the Internal Cartridge Wash 1 step is conducted.
The Collect Flow Through step elutes the wash flow-through from the Internal Cartridge Wash 1 step directly into the Flow Through Collection plate (deck location 7). If the Internal Cartridge Wash step is conducted, but the Collect Flow Through step is not selected, the wash flow-through is dispensed directly into the Waste plate (deck location 3).
For the maximum practical working volumes per well for specific labware at deck location 7, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
Select this step if you want to collect the flow-through generated during the Internal Cartridge Wash 1 step. In the default protocol, this step is not selected.
Load Metal Reagent step
The Load Metal Reagent step aspirates the specified volume of Metal Reagent (deck location 9) into the syringes, performs an external syringe wash at the wash station (deck location 1) to remove any Metal Reagent remaining on the outside of the probes, and then mounts the cartridges at the seating station (deck location 2). The Metal Reagent is dispensed through the cartridges at the specified flow rate into the Waste plate (deck location 3), or into the Flow Through Collection plate (deck location 7) if the Collect Flow Through step is selected.
An external cartridge tip wash is performed at the wash station (deck location 1), where the cartridge tips dip into the chimneys to remove any Metal Reagent on the outside of the cartridge. After removing the cartridges at the seating station (deck location 2), the syringes are washed the specified number of cycles at the wash station (deck location 1) using 250 µL per wash cycle.
Always select the Load Metal Reagent step during a normal run. In the default protocol, this step is selected.
Table Setting the Load Metal Reagent step parameters
 
Collect Flow Through (Load Metal Reagent step)
Note: The Collect Flow Through step is an option only if the Load Metal Reagent step is conducted.
If you select the Collect Flow Through step, the flow-through from the Load Metal Reagent step is dispensed directly into the Flow Through Collection plate (deck location 7).
If the Load Metal Reagent step is conducted, but the Collect Flow Through step is not selected, the flow-through is dispensed directly into the Waste plate (deck location 3).
For the maximum practical working volumes per well for specific labware at deck location 7, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
Select the Collect Flow Through step if you want to collect the flow-through generated during the Load Metal Reagent step for downstream processing or for analysis. In the default protocol, this step is not selected.
Cup Wash 2 step
The Cup Wash 2 step removes the small volume of residual liquid that might remain above the upper cartridge membrane after the Load Metal Reagent step. The Cup Wash 2 step effectively rinses the cup portion of AssayMAP cartridges with buffer. The Cartridge Wash Buffer 2 (deck location 8) is aspirated into the syringes, and then dispensed into the cartridge cups (deck location 2). This liquid, plus any residual liquid, is aspirated from the cartridge cups into the syringes. The protocol ensures that no cartridges are stuck to the probes before dispensing the liquid into the Waste plate (deck location 3), and then washing the syringes at the wash station (deck location 1).
Select the Cup Wash 2 step if residual liquid remains in the cartridge cups from the Load Metal Reagent step. In the default protocol, this step is selected.
Table Setting the Cup Wash 2 step parameters
 
Internal Cartridge Wash 2 step
The Internal Cartridge Wash 2 step aspirates Cartridge Wash Buffer 2 (deck location 8) into the syringes. After mounting the cartridges at the seating station (deck location 2), the buffer is dispensed through the cartridges at the specified flow rate into the Waste plate (deck location 3), or into the Flow Through Collection plate (deck location 7) if the Collect Flow Through step is selected. An external cartridge tip wash is performed at the wash station (deck location 1), where the cartridge tips dip into the chimneys to remove any Cartridge Wash Buffer 2 on the outside of the cartridge. After removing the cartridges at the seating station (deck location 2), the syringes are washed the specified number of cycles at the wash station (deck location 1) using 250 µL per wash cycle.
Select the Internal Cartridge Wash 2 step to flush away any remaining Metal Reagent and prepare the cartridge with a buffer compatible with immediate use or short-term storage. In the default protocol, this step is selected.
Table Setting the Internal Cartridge Wash 2 step parameters
 
Collect Flow Through step (from the Internal Cartridge Wash 2 step)
Note: The Collect Flow Through step is an option only if the Internal Cartridge Wash 2 step is conducted.
The Collect Flow Through step dispenses the wash flow-through from the Internal Cartridge Wash 2 step directly into the Flow Through Collection plate (deck location 7). If the Internal Cartridge Wash 2 step is conducted, but the Collect Flow Through step is not selected, the wash flow-through is dispensed directly into the Waste plate (deck location 3).
For the maximum practical working volumes per well for specific labware at deck location 7, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
Select this step if you want to collect the flow-through generated during the Internal Cartridge Wash 2 step. In the default protocol, this step is not selected.
Final Syringe Wash step
The Final Syringe Wash step uses the wash station (deck location 1) to flush any potential contaminants from the syringes in the head. During each Final Syringe Wash cycle, the head aspirates 250 µL into the syringes from the wash station chimneys, and then moves by a fixed offset between the chimneys to dispense the syringe contents to waste.
Select the Final Syringe Wash step after each run to clean the syringes thoroughly before running a different assay. In the default protocol, this step is selected.
Table Setting the Final Syringe Wash parameter
 
The amount of carryover in the syringe (expressed as dilution of the starting sample) depends on the total number of syringe wash cycles. Three cycles give a dilution of up to 103. In cases where carryover is a major concern, increasing the number of iterations also increases the dilution factor, but with a cost of increased assay time and reduced syringe lifetime.
 
Return to top
Automation movements during the protocol
This section describes the basic automation movements of the AssayMAP Bravo Platform during the protocol when the protocol defaults are used. Changing the selections or deselecting options will alter the movements. The default protocol takes approximately 60 minutes to complete.
 
Parks the cartridges in column A of the 96AM Cartridge & Tip Seating Station.
Return to top