• Greater than the actual number of columns used, the Bravo Platform will apply too much force when mounting the cartridges, which can damage both the cartridges and the AssayMAP syringes in the head.
• Less than the actual number of columns used, the Bravo Platform will not apply enough force to seat the cartridges properly. Volume (µL). The default volume should be sufficient to wet the cartridge resin and remove entrained air from within the cartridge. Using less than the default volume may leave air in the resin bed. Using more than the default volume is unnecessary and increases run time.Flow rate (µL/min). A flow rate slower than the default value diminishes the ability to effectively remove entrained air from the cartridge. A flow rate faster than the default is not required and has not been tested.Wash cycles. 250 µL of DI water is used for each syringe wash cycle. Volume (µL). The default volume is equal to 10 column volumes, which should be sufficient for complete buffer exchange. Using less than the default volume may not fully equilibrate the resin bed. Using more than the default volume is unnecessary and increases run time.Flow rate (µL/min). A flow rate slower than the default rate, 10 µL/min, will likely have no benefit, but will increase the total assay time. A flow rate faster than 20 µL/min might not equilibrate through the pores in the beads.Wash cycles. 250 µL of DI water is used for each syringe wash cycle.
• # of times to load = total sample volume/250,
where # times to load is rounded up to nearest integer
• volume of each load = sample volume/# of times to load# times to load = 900/250 = 3.6, which is rounded up to 4volume of each load = 900/4 = 225If Collect Flow Through is selected for the Load Samples step, be sure that the Flow Through Collection plate has sufficient maximum well capacity. For details, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench. Flow rate (µL/min). The optimum sample loading flow rate requires balancing the speed of the assay and desired recovery. When setting the flow rate, be aware that the quantitative binding capacity is inversely proportional to the flow rate. For some samples, a flow rate faster than 15 µL/min results in a loss of total binding capacity in addition to a loss of quantitative capacity, as the sample does not have sufficient time to diffuse throughout the pores of the cartridge resin bed.Wash cycles. 250 µL of DI water is used for each syringe wash cycle. Volume (µL). A volume <10 µL might be insufficient for cup washing, while a volume >50 µL may offer little benefit.Wash cycle. Each cycle comprises one cup wash and one syringe wash. Volume (µL). The volume to dispense through the cartridges to wash the resin bed. Volumes higher than the default volume may improve the purification marginally but also increase the run time. Volumes lower than the default volume may be insufficient for efficient cartridge washing.Note: Setting the volume to zero skips all Internal Cartridge Wash tasks except for syringe washing.Flow rate (µL/min). A rate slower than the default flow rate, 10 µL/min, will likely have little benefit, but will increase the total assay time. A rate faster than 20 µL/min might not equilibrate through the pores in the beads, resulting in incomplete washing.Wash cycle. The number of syringe wash cycles to perform at the end of the task. 250 µL of DI water is used for each syringe wash cycle. Volume (µL). A volume <10 µL may be insufficient for cup washing, while a volume >50 µL may offer little benefit.Wash cycle. Each cycle comprises one cup wash and one syringe wash. Volume (µL). The volume to dispense through cartridges to wash the resin bed. Volumes higher than the default volume may improve the purification marginally but will also increase the run time. Volumes lower than the default volume may be insufficient for efficient cartridge washing.Flow rate (µL/min). A rate slower than the default flow rate, 10 µL/min, will likely have little benefit, but will increase the total assay time. A rate faster than 20 µL/min might not equilibrate through the pores in the beads, resulting in incomplete washing.Wash cycle. The number of syringe washes to perform at the end of the task after the cartridges are parked in the seating station. 250 µL of DI water is used for each syringe wash cycle. Volume (µL). The volume of Elution & Syringe Wash Buffer to aspirate and dispense. Volumes higher than the default volume are unlikely to improve the syringe cleaning but will increase the run time. Volumes lower than the default volume may be insufficient for efficient syringe washing.Note: Setting the volume to zero will skip all tasks associated with the Stringent Syringe Wash step.Wash cycle. A wash cycle is a stringent syringe wash followed by a basic syringe wash at the wash station. Volume (µL). The maximum volume of eluate that can be collected in the Eluate Collection plate is dependent on the maximum practical working volume of the labware defined for the Affinity Purification protocol. The total volume in the Eluate Collection plate is determined by summing the net elution volume (Elute volume - Eluate Discard volume) and the Existing Collection Volume. For labware-specific maximum well volumes, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.Flow rate (µL/min). A flow rate slower than 5 µL/min is unlikely to improve the elution yield. Elution yield may be compromised if flow rates are faster than 15 µL/min for a given volume of elution buffer (that is, more elution buffer may be required to get the same elution yield at high elution flow rates relative to using lower flow rates for a given elution volume).Wash cycle. The number of syringe washes to perform at the wash station after an Elute step. 250 µL of DI water is used for each syringe wash cycle. Example: If the Elute, Eluate Discard, and Add to Flow Through steps are all selected with the following settings,For AssayMAP cartridges, the initial elution volume (~2 µL for 5 µL cartridges and ~10 µL for the 25 µL cartridges) contains small or no measurable amounts of analyte. The volume of the analyte-free eluate will vary depending on both the strength of the elution buffer and the degree of analyte saturation of the cartridges. This value should be determined empirically for each combination of sample, buffer, cartridge, and system.Volume (µL). The first volume of eluate that will be discarded during the Elute step. This value can equal, but cannot exceed the Elute volume. For the maximum practical working volumes of labware for eluate collection, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench. This section describes the basic movements of the AssayMAP Bravo Platform during the Affinity Purification protocol using the default protocol settings. Changing the selections or parameters will alter the movements.