|
• Greater than the actual number of used, the Bravo Platform will apply too much force when mounting the cartridges, which can cause damage to both the cartridges and the AssayMAP syringes in the head.
• Less than the actual number used, the Bravo Platform will not apply enough force to seat the cartridges properly. Volume (µL). The default volume is equal to 10 column volumes, which should be sufficient for complete buffer exchange. Using less than the default volume may not fully equilibrate the resin bed. Using more than the default volume is unnecessary and increases run time.Flow rate (µL/min). A flow rate slower than 10 µL/min, will likely have no benefit, but will increase the total assay time. A flow rate faster than 20 µL/min might not equilibrate through the pores in the beads. # of times to load = total sample volume/250,
where # times to load is rounded up to nearest integervolume of each load = sample volume/# of times to load# times to load = 900/250 = 3.6, which is rounded up to 4volume of each load = 900/4 = 225If Collect Flow Through is selected for the Load Samples step, be sure that the Flow Through Collection plate has sufficient maximum well capacity. For details, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.IMPORTANT The volume of sample to load should be balanced with the sample concentration and the mass capacity of the cartridge. Be sure to include the recommended labware-dependent volume overage to prevent air from entering the cartridge. For more information, see Preparing the sample plates.Volume (µL). The volume of sample should be balanced with the sample concentration and the mass capacity of the cartridge. To determine the volume of sample to load, see Determining the volume of sample to load. Flow rate (µL/min). The optimum sample loading flow rate requires balancing the speed of the assay and desired recovery. When setting the flow rate, be aware that the quantitative binding capacity is inversely proportional to the flow rate. For some samples, a flow rate faster than 15 µL/min can result in a low dynamic binding capacity. It might also produce some loss of total binding capacity, as the sample does not have sufficient time to diffuse throughout the pores of the cartridge resin bed.Wash cycles. 250 µL of DI water is used for each syringe wash cycle. Volume (µL). A volume <10 µL might be insufficient for cup washing, while a volume >50 µL may offer little benefit.Wash cycle. Each cycle comprises one cup wash and one syringe wash. Volume (µL). The volume to dispense through the cartridges to wash the resin bed. Volumes higher than the default volume may improve the purification marginally but will also increase the run time. Volumes lower than the default volume may be insufficient for efficient cartridge washing.Flow rate (µL/min). A rate slower than 10 µL/min will have little benefit, but will increase the total assay time. A rate faster than 20 µL/min might not equilibrate through the pores in the beads, resulting in incomplete washing.Wash cycle. The number of syringe wash cycles to perform at the end of the task. 250 µL of DI water is used for each syringe wash cycle. Volume (µL). A volume <10 µL may be insufficient for cup washing, while a volume >50 µL may offer little benefit.Wash cycle. Each cycle comprises one cup wash and one syringe wash. Volume (µL). The volume to dispense through cartridges to wash the resin bed. Volumes higher than the default volume may improve the purification marginally but will also increase the run time. Volumes lower than the default volume may be insufficient for efficient cartridge washing.Flow rate (µL/min). A rate slower than 10 µL/min will have little benefit, but will increase the total assay time. A rate faster than 20 µL/min might not equilibrate through the pores in the beads, resulting in incomplete washing.Wash cycle. The number of syringe washes to perform at the end of the task after the cartridges are parked in the seating station. 250 µL of DI water is used for each syringe wash cycle. Volume (µL/min). The volume of Elution & Syringe Wash Buffer to aspirate and dispense. Volumes higher than the default volume are unlikely to improve the syringe cleaning but increase run time. Volumes lower than the default volume may be insufficient for efficient syringe washing.Note: Setting the volume to zero will skip all tasks associated with the Stringent Syringe Wash step.Wash cycle. A wash cycle is a stringent syringe wash followed by a basic syringe wash at the wash station. Volume (µL). Using sufficiently strong elution conditions and the default flow rate (5 µL/min), targets can typically be fully eluted from cartridges in approximately two cartridge column volumes.Flow rate (µL/min). A flow rate slower than 5 µL/min is unlikely to improve the elution yield. Elution yield may be compromised if flow rates are faster than 15 µL/min for a given volume of elution buffer (that is, more elution buffer may be required to get the same elution yield at high elution flow rates relative to using lower flow rates for a given elution volume). Volume (µL). The default volume is equal to 10-column volumes, which should be sufficient for complete buffer exchange. Using less than the default volume may not fully equilibrate the resin bed. Using more than the default volume is unnecessary and increases run time.Flow rate (µL/min). A flow rate slower than 10 µL/min, will likely have no benefit, but will increase the total assay time. A flow rate faster than 20 µL/min might not equilibrate through the pores in the beads across the full length of the cartridge bed. Note: The amount of carryover in the syringe (expressed as dilution of the starting sample) depends on the total number of syringe wash cycles. Three cycles give a dilution of up to 103. In cases where carryover is a major concern, increasing the number of iterations also increases the dilution factor, but with a cost of increased assay time and reduced syringe lifetime.
• Load Samples: Set the number of wash cycles to 0.
• Collect Flow Through: Not selected.
• Cup Wash 1: Not selected.
• Internal Cartridge Wash 1: Set the number of wash cycles to 0. (If Internal Cartridge Wash 2 is not selected, set the number of wash cycles to any desired value.)
• Cup Wash 2: Not selected.This section describes the basic movements of the AssayMAP Bravo Platform during the Affinity Purification: Aspiration Mode protocol using the default protocol settings. Changing the selections or parameters will alter the movements.