AssayMAP Protein Sample Prep Workbench : Fractionation v1.1 User Guide : Before you start

Before you start
This topic lists the required hardware, software, AssayMAP kits, labware, and reagents for running the Fractionation protocol. If you have questions about these items, contact Agilent Customer Service.
Hardware
Required hardware
 
 
The following figure and table show the components of the AssayMAP Bravo Platform, which are required for running the AssayMAP protocols.
Figure. AssayMAP Bravo Platform components
 
 
96AM Wash Station or the later model 96 Channel Wash Station
Pump Module 2.0 and two carboys
10 
Orbital Shaking Station with Control Unit
 
*To avoid a hardware crash and equipment damage, ensure that the wash station contains the white wide-bore chimneys when using the 25 µL cartridges.
 
Note: The white wide-bore chimneys work for both 5-µL and 25-µL cartridges and are standard on wash stations purchased in 2020 onward. The wide-bore chimneys are white plastic, whereas the standard-bore chimneys are a semi-clear plastic. For details, see the 96 Channel Wash Station Maintenance Guide.
 
 
Optional equipment
You might need the following when preparing the samples and reagents:
Microplate centrifuge, such as the Agilent Microplate Centrifuge or equivalent
Microplate sealer, such as the Agilent PlateLoc Thermal Microplate Sealer or equivalent
 
Return to top
Software
 
 
The following table lists the minimum software requirements:
 
Agilent VWorks Automation Control
Agilent Protein Sample Prep Workbench
 
 
Return to top
Starter kit, cartridges, and labware
The following table displays the starter kit that is available for the Fractionation application. The starter kit contains both strong cation exchange (SCX) cartridges and labware.
Note: There is no Fractionation starter kit that contains C18 or reversed-phase (RP-S) cartridges. To order C18 or RP-S cartridges, see AssayMAP cartridges.
 
AssayMAP cartridges
You can purchase each of the following cartridge types as an individual rack of 96 cartridges without the labware. To order the cartridges, contact Automation Solutions Customer Service.
Note: The SCX cartridges are included in the starter kit.
Note: This application has not yet been optimized for the 25 µL cartridges.
 
Cartridge use and storage guidelines
See the cartridge box label for storage guidelines.
Follow these guidelines to get the best performance from AssayMAP cartridges:
 
*Cartridges ship dry and, therefore, contain air entrained in the cartridge bed. Failure to prime the cartridges can prevent the reagents and buffers from accessing parts of the resin bed, resulting in reduced capacity and poor reproducibility.
 
Note: Cartridges will not dry out during the course of a normal application run. Cartridges can dry out rapidly if they are exposed to air for extended periods (e.g., >1 hour) after they have been primed and equilibrated.
If you need to store primed and equilibrated cartridges for a short period, ensure that you use the lidded blue rack-receiver plate stack with an appropriate solution in the receiver plate chimneys such that the cartridge tips are submerged in the solution.
Labware
 
*Use only the labware supplied with the starter kit and place them at the AssayMAP Bravo deck locations specified in the instructions. Using other labware or placing labware at unapproved deck locations can cause crashes during protocol runs and damage hardware components.
 
The following labware are required for the Fractionation application. They are included in the starter kit. However, you can also purchase them separately.
 
*For dimensionally equivalent alternatives and other details about the labware, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
 
Return to top
Reagents
The AssayMAP Fractionation protocol is designed to permit multiple types of fractionation to be performed. Therefore, the reagent requirements are dependent on the fractionation strategy and the chemistry of the AssayMAP cartridges.
In general, buffer systems used for liquid chromatography separations in a column format can be adapted for use in the AssayMAP cartridge format. Fractionation using AssayMAP SCX cartridges can be accomplished using elution buffers with increasing ionic strength. Alternatively, fractionation can be achieved using elution buffers with increasing pH and SCX cartridges.
Similarly, both C18 and RP-S cartridges can be used to for low-pH, reversed-phase fractionation using increasing concentrations of organic solvent in the elution buffers. Because of the high-pH stability of RP-S cartridges, they may be used to achieve high-pH, reversed-phase fractionation with increasing concentrations of organic solvent.
 
Return to top