|
• Greater than the actual number of columns used, the Bravo Platform will apply too much force when mounting the cartridges, which can damage both the cartridges and the AssayMAP syringes in the head.
• Less than the actual number of columns used, the Bravo Platform will not apply enough force to seat the cartridges properly. Volume (µL). The default volume should be sufficient to wet the cartridge resin and remove entrained air from within the cartridge. Using less than the default volume may leave air in the resin bed. Using more than the default volume is unnecessary and increases run time.Flow rate (µL/min). A flow rate slower than the default value diminishes the ability to effectively remove entrained air from the cartridge. A flow rate faster than the default is not required and has not been tested.Wash cycles. 250 µL of DI water is used for each syringe wash cycle. Volume (µL). The default volume is equal to 10 column volumes, which should be sufficient for complete buffer exchange. Using less than the default volume may not fully equilibrate the resin bed. Using more than the default volume is unnecessary and increases run time.Flow rate (µL/min). A flow rate slower than the default rate will likely have no benefit, but will increase the total assay time. A flow rate faster than 20 µL/min might not equilibrate through the pores in the beads.Wash cycles. 250 µL of DI water is used for each syringe wash cycle. The protocol accommodates sample volumes up to 1000 µL to be dispensed through the cartridges. Although, the form permits you to enter smaller volumes, the minimum advisable sample volume to be loaded onto an AssayMAP cartridge is 10 µL.
• # of times to load = total sample volume/250,
where # times to load is rounded up to nearest integer
• volume of each load = sample volume/# of times to load# times to load = 900/250 = 3.6, which is rounded up to 4volume of each load = 900/4 = 225If Collect Flow Through is selected for the Load Samples step, be sure that the Flow Through Collection plate has sufficient maximum well capacity. For details, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench. Flow rate (µL/min). The optimum sample loading flow rate requires balancing the speed of the assay and desired recovery. When setting the flow rate, be aware that the quantitative binding capacity is inversely proportional to the flow rate. For some samples, a flow rate faster than 15 µL/min can result in a loss of total binding capacity in addition to a loss of quantitative capacity, as the sample does not have sufficient time to diffuse throughout the pores of the cartridge resin bed.Wash cycles. 250 µL of DI water is used for each syringe wash cycle. Volume (µL). A volume <10 µL might be insufficient for cup washing, while a volume >50 µL may offer little benefit.Wash cycle. Each cycle comprises one cup wash and one syringe wash. Volume (µL). The volume to dispense through the cartridges to wash the resin bed. Volumes higher than the default volume may improve the purification marginally but also increase the run time. Volumes lower than the default volume may be insufficient for efficient cartridge washing.Flow rate (µL/min). A rate slower than the default flow rate, 10 µL/min, will have little benefit, but will increase the total assay time. A rate faster than 20 µL/min might not equilibrate through the pores in the beads, resulting in incomplete washing.Wash cycle. The number of syringe wash cycles to perform at the end of the task. 250 µL of DI water is used for each syringe wash cycle. Volume (µL). The volume of Blocking Reagent to aspirate into the syringes and dispense through the cartridges.Flow rate (µL/min). A rate slower than 5 µL/min will likely have no benefit, but will increase the total assay time. A rate faster than 20 µL/min might not equilibrate through the pores in the beads, which may result in incomplete blocking.Wash cycles. 250 µL of DI water is used for each syringe wash cycle. Volume (µL). A volume <10 µL may be insufficient for cup washing, while a volume >50 µL may offer little benefit.Wash cycle. Each cycle comprises one cup wash and one syringe wash. Volume (µL). The volume to dispense through cartridges to wash the resin bed.Flow rate (µL/min). A rate slower than the default flow rate, 10 µL/min, will have little benefit, but will increase the total assay time. A rate faster than 20 µL/min might not equilibrate through the pores in the beads, resulting in incomplete washing.Wash cycle. The number of syringe washes to perform at the end of the task after the cartridges are parked in the seating station. 250 µL of DI water is used for each syringe wash cycle. If the Collect Flow Through step is not selected, the flow-through is dispensed at the wash station. Volume (µL). The volume of Stringent Syringe Wash Buffer to aspirate and dispense. Volumes higher than the default volume are unlikely to improve the syringe cleaning but will increase the run time. Volumes lower than the default volume may be insufficient for efficient syringe washing.Wash cycle. A wash cycle is a stringent syringe wash followed by a basic syringe wash at the wash station. Volume (µL). The default volume is equal to 10 column volumes, which should be sufficient for complete buffer exchange.Flow rate (µL/min). A flow rate slower than the default rate will likely have no benefit, but will increase the total assay time. A flow rate faster than 20 µL/min might not equilibrate through the pores in the beads in the cartridge resin bed.Wash cycles. 250 µL of DI water is used for each syringe wash cycle. This section describes the basic movements of the AssayMAP Bravo Platform during the Immobilization protocol. Changing the selections or parameters will alter the movements.