AssayMAP Protein Sample Prep Workbench : In‑Solution Digestion: Multi-Plate v1.1 User Guide : Preparing the reagents

Preparing the reagents
The In-Solution Digestion protocol accommodates a wide range of digestion chemistries so that previously optimized conditions can be easily transferred to this automated platform. Therefore the reagents described in this topic are categories of reagents used in the assay rather than specific recommendations.
The following reagents are required for the In-Solution Digestion protocol:
 
Use the Single Liquid Addition v1.1 utility to transfer 300 µL per well of the wash solution to the four Greiner 96-Well Round Bottom, White Plates that will be stacked on deck location 2 prior to running the In-Solution Digestion protocol. During the protocol run, these plates serve two purposes. The plates are moved on top of the sample plates and serve as lids to protect the samples from light during the alkylation step of the protocol. The plates are also used as reservoirs for the syringe wash solution which is used to wash the syringes to prevent carry-over between sample loading steps.
Use the supplied Reagent Volume Calculator to determine the optimal reagent formulations. Run the Reagent Plate Setup protocol using the calculated values to dispense the reagents into the appropriate labware.
 
*A small reagent volume excess is required in all labware types to ensure proper volume transfer. Use the Reagent Volume Calculator to automatically include excess volume, or look up the recommended values for each labware type in the Labware Reference Guide.
 
Note: You can find the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
About the Reagent Volume Calculator for In-Solution Digestion
The Reagent Volume Calculator is a Microsoft Excel worksheet that:
The calculator consists of the following worksheets:
 
Return to top
Opening the Reagent Volume Calculator for In-Solution Digestion: Multi-Plate
To open the Reagent Volume Calculator:
1
2
Locate In-Solution Digestion: Multi-Plate, and then click Calculator. Microsoft Excel starts and displays the calculator.
3
 
Return to top
Entering values in the Reaction Setup worksheet
The Reaction Setup worksheet consists of three tables:
Based on the values you supply in the green boxes, the software will calculate and update the read-only values (gray boxes) in the rest of the worksheet.
The boxes for which you are allowed to enter values should remain green after you enter a value. If you enter a value that meets only some of the requirements, the box becomes yellow. If you enter a value that does not meet any requirement, the box becomes red.
To enter values in the Reaction Setup worksheet:
1
In the Reaction Setup table, type the values in the green boxes in the 1. Sample Input area:
 
 
Read only. The calculator converts the Number of Samples to the Total Number of Sample Columns by rounding up to the nearest multiple of 8.
Read only. The number of columns that occupy each sample plate. The value is calculated using the Total Number of Sample Columns and the Number of Sample Plates.
Read only. The mass of the sample. The value is calculated using the Sample Volume and Protein Concentration.
2
In the Reaction Setup table, type the values in the green boxes in the 2. Denaturation/Reduction area:
 
 
Read only. The concentration of guanidine or urea in the Master Reagent.
Read only. The micromoles of reductant. The value is calculated using the Reductant Concentration and the Denaturation Mixture volume.
Optional. The concentration of the internal standard stock solution you want to add to the Denaturation Mixture.
3
In the Reaction Setup table, type the values in the green boxes in the 3. Alkylation/Reaction area:
 
 
Read only. The number of micromoles of the alkylant. The value is calculated from the concentration and volume added to the reaction, and the final concentration at the indicated steps is displayed in the B. Reactant Concentrations Summary table.
4
In the Reaction Setup table, type the values in the green boxes in the 4. Dilution/Alkylation Quench area:
 
 
Read only. The number of micromoles of reductant that is added during this step.
Read only. The volume of the Diluent Mixture you want to add.
5
In the Reaction Setup table, type the values in the green boxes in the 5. Digestion area:
 
 
Read only. The volume of Protease to add.
Read only. The final concentration of the buffer.
Read only. The digest reaction volume.
6
Review the information in the B. Reactant Concentration Summary table. If you need to make any adjustments, repeat steps 1 through 5.
 
7
Review the information in the C. In-Solution Digest form inputs table. You will transfer these values when setting up the In-Solution Digestion protocol. See Running the protocol.
 
 
Return to top
Entering values in the Reagent Prep worksheet
The Reaction Prep worksheet consists of four tables:
Each table displays the recipe for the named reagent. Notice that the tables are prepopulated with values that have been calculated using your input in the Reaction Setup worksheet.
If you are using more denaturant than what is required, you can enter the larger mass in the Denaturation Mixer table.
The Master Reagent requires iodoacetamide powder. If you are using more than what is required, you can enter the larger mass in the Alkylant table.
 
 
Return to top
Dispensing the reagents
 
*To minimize evaporation, fill the labware immediately before run time or keep the plates lidded until the run begins.
 
Before you start, make sure you have the required labware. See Labware and starter kits.
Prepare the reagents according to the recipes shown in the Reagent Prep worksheet of the Reagent Volume Calculator.
To dispense the reagents into the reservoirs:
1
2
In the Reagent Volume Calculator, do one of the following:
Automated Reagent Plate Setup. In the Reagent Volume Calculator, display the Automated Plate Setup worksheet, and then proceed to step 3.
Manual Reagent Plate Setup. In the Reagent Volume Calculator, display the Manual Plate Setup worksheet. Use manual pipettes to prepare reagent plates for the Protease, Alkylant, Denaturation, and Diluent reagents based on their respective plate layouts. Then proceed to preparing the samples (Preparing the samples) and running the protocol (Running the protocol).
3
 
4
 
5
Prepare the Syringe Wash Buffer plate. Use the Single Liquid Addition v1.1 utility to fill four Greiner 96-Well Round Bottom, White Plates with the Syringe Wash Buffer.
6
7
 
Return to top
Running the Reagent Plate Setup protocol
The Reagent Plate Setup protocol uses the AssayMAP Bravo Platform to transfer reagents in the Master Reagent plate to dedicated reagent plates.
To run the protocol:
1
2
Locate In-Solution Digestion: Multi-Plate, and then click App.
 
The In-Solution Digestion: Multi-Plate app opens.
 
3
 
4
In the Application Settings area, specify the following:
 
5
In the Step table:
a
b
 
*Use the values shown in the Reagent Volume Calculator, Reaction Setup worksheet, table C. In-Solution Digest form inputs. The total volume of Denaturation Mixture, Alkylant, Protease, Diluent Mixture, and sample must be less than 300 µL to accommodate the working volume of the U-bottom labware. The total volume should also include any liquid that will be added to the sample plate after digestion (for example, quenching by acidification).
6
Optional. Click Save Settings.
7
Confirm that the physical layout on the AssayMAP Bravo deck matches the Labware Table area and the Deck Layout area of the application.
 
*Incorrect labware can cause a hardware collision, resulting in equipment damage. Ensure that the physical labware present on the Bravo deck exactly match the Labware Table and Deck Layout areas of the application.
 
The tip box at deck location 6 must contain at least three full columns of pipette tips. If you have a partially filled tip box, make sure the empty columns are on the right side of the tip box.
8
 
*Improperly seated labware can cause a hardware collision, resulting in equipment damage. Ensure that all labware are properly seated within the alignment features of their respective platepads.
 
9
Click Run Protocol. The protocol run starts.
To monitor the progress of the run, check the Status box.
 
*To stop a run in an emergency, use the hardware Emergency Stop button.
 
To pause the run, click Pause. The task currently in progress finishes before the protocol pauses. The Scheduler Paused dialog box opens. For details, see Emergency stops and pauses.
 
Return to top
Cleaning up
When the protocol run is finished, make sure you:
Remove the 96AM Cartridge & Tip Seating Station from the Bravo deck and discard used pipette tips.
 
*Make sure you discard the chemical waste and used labware according to your lab’s waste disposal procedures and in compliance with all local, state, and federal safety regulations.
 
 
Return to top