AssayMAP Protein Sample Prep Workbench : Peptide Cleanup: Aspiration Mode v2.0 User Guide : Assay development guidelines and protocol notes

Assay development guidelines and protocol notes
This topic explains each step of the Peptide Cleanup: Aspiration Mode protocol so that you can optimize the protocol to your particular experimental design.
Protocol stepwise guidelines
 
Greater than the actual number of columns used, the Bravo Platform will apply too much force when mounting the cartridges, which can damage both the cartridges and the AssayMAP syringes in the head.
Less than the actual number of columns used, the Bravo Platform will not apply enough force to seat the cartridges properly.
Volume (µL). The default volume is equal to 10-column volumes, which should be sufficient for complete buffer exchange. Using less than the default volume may not fully equilibrate the resin bed. Using more than the default volume is unnecessary and increases run time.
Flow Rate (µL/min). A flow rate slower than 10 µL/min will likely have no benefit, but will increase the total assay time. A flow rate faster than 20 µL/min might not equilibrate through the pores in the beads across the full length of the cartridge bed.
This step mounts the cartridges on the syringes, aspirates the samples through the cartridges, and performs an external cartridge tip wash at the wash station to remove any sample on the outside of the cartridge. This wash is followed by aspirating 5 µL of Equilibration Buffer (sample chase) through the cartridges at the Load Samples flow rate. The sample chase ensures that the sample volume in the cartridges at the end of the sample load moves through the cartridge bed at the same rate as the rest of the sample. Finally, the cartridges are parked at the seating station and the flow-through plus sample chase are collected in the Flow Through Collection plate.
Volume (µL). The volume of sample should be balanced with the sample concentration and the mass capacity of the cartridge. To determine the volume of sample to load, see Determining the volume of sample to load.
Flow rate (µL/min). The optimum sample loading flow rate requires balancing the speed of the assay and desired recovery. When setting the flow rate, be aware that the quantitative binding capacity is inversely proportional to the flow rate. For some samples, a flow rate greater than 15 µL/min can result in a low dynamic binding capacity. It might also produce some loss of total binding capacity, as the sample does not have sufficient time to diffuse throughout the pores of the cartridge resin bed. Using flow rates below 5 µL/min might not significantly increase analyte binding.
Wash Cycles. 250 µL of DI water is used for each syringe wash cycle.
Volume (µL). A volume <10 µL might be insufficient for cup washing, while a volume >50 µL may offer little benefit.
Volume (µL). The volume to dispense through the cartridges to wash the resin bed. Volumes higher than the default volume may improve the purification marginally but also increase the run time. Volumes lower than the default volume may be insufficient for efficient cartridge washing.
Flow Rate (µL/min). A rate slower than 10 μL/min will likely have no benefit, but will increase the total assay time. A rate faster than 20 μL/min might not equilibrate through the pores in the beads, resulting in incomplete washing.
Wash Cycles. The number of syringe washes to perform at the end of the task. 250 µL of DI water is used for each syringe wash cycle.
Volume (µL): The volume of Priming & Stringent Syringe Wash Buffer to aspirate and dispense. Volumes lower than the default volume may be insufficient for efficient syringe washing.
Wash Cycles. A wash cycle is a stringent syringe wash followed by a basic syringe wash at the wash station.
Volume (µL). Using sufficiently strong elution conditions and the default flow rate (5 µL/min), peptides can be fully eluted from cartridges in approximately two cartridge column volumes.
Flow Rate (µL/min). A flow rate slower than 5 µL/min is unlikely to improve the elution yield. Elution yield may be compromised if flow rates are faster than 15 µL/min for a given volume of elution buffer (that is, more buffer may be required to get the same elution yield at high elution flow rates relative to using lower flow rates for a given elution volume).
Wash Cycles. The number of syringe washes to perform at the wash station after an Elute step. 250 ?L of DI water is used for each syringe wash cycle.
Volume (µL). The default volume is equal to 10-column volumes, which should be sufficient for complete buffer exchange. Using less than the default volume may not fully equilibrate the resin bed. Using more than the default volume is unnecessary and increases run time.
Flow Rate (µL/min). A flow rate slower than 10 µL/min will likely have no benefit, but will increase the total assay time. A flow rate faster than 20 µL/min using the default volume might not equilibrate through the pores in the beads.
 
Return to top
Automation movements during the protocol
This section describes the basic movements of the AssayMAP Bravo Platform during the Peptide Cleanup: Aspiration Mode protocol using the default settings. Changing the selections or parameters will alter the movements.
 
 
Return to top