AssayMAP Protein Sample Prep Workbench : Phosphopeptide Enrichment v2.1 User Guide : Assay development guidelines and protocol notes

Assay development guidelines and protocol notes
This topic explains each step of the protocol so that you can optimize the Phosphopeptide Enrichment protocol to your particular experimental design.
Note: For guidelines on performing a mock run, see About performing a mock run (optional).
Protocol stepwise guidelines
 
*Cartridges ship dry and, therefore, contain air entrained in the cartridge bed. Failure to prime the cartridges can prevent the sample and buffers from accessing parts of the resin bed, resulting in reduced capacity and poor reproducibility.
 
The first steps in the Phosphopeptide Enrichment protocol set up the system and prepare the cartridges for sample loading, including:
 
*Do not allow wetted cartridges to dry out.
 
Number of Full Columns of Cartridges
This setting is critical to set the proper force used to mount the cartridges. To obtain expected instrument performance, ensure that the column selection is correct. If the column selection is:
Greater than the actual number of columns used, the Bravo Platform will apply too much force when mounting the cartridges, which can cause damage to both the cartridges and the AssayMAP syringes in the head.
For example, if the software specifies 10 columns, but only 1 column of cartridges are in the seating station, the head will apply 12 times more force than what is required.
Less than the actual number of columns used, the Bravo Platform will not apply enough force to seat the cartridges properly.
For example, if the software specifies 1 column, but 12 columns of cartridges are in the seating station, the head will apply 1/12th the force required to seat the cartridges properly. In this case, cartridges may fall off during the run or the volume of liquid that moves across the cartridge bed may be variable do to liquid moving past the syringe cartridges seal into the cartridge cup.
Default: 1
Range: 1-12
Initial Syringe Wash step
The Initial Syringe Wash step removes any potential contaminants from the syringes. During each Initial Syringe Wash cycle, the head aspirates 250 µL into the syringes from the wash station chimneys, and then moves by a fixed offset between the chimneys to dispense the syringe contents to waste.
Select the Initial Syringe Wash step to clean the syringes. Skip this step if you are certain that the syringes are clean. This step is selected by default.
Table Setting the Initial Syringe Wash parameter
 
Prime step
The Prime step removes entrained air from within the cartridges and properly wets the surface of the resin. For the most effective priming of AssayMAP Fe(III)-NTA and TiO2 cartridges, the Priming Buffer requires that the solution contain at least 50% organic solvent. Higher concentrations of organic solvent are also acceptable.
The Prime step aspirates the specified volume of Priming Buffer (deck location 5) into the syringes, mounts the cartridges (deck location 2) onto the syringes, and then dispenses buffer at the specified flow rate through the cartridges into the Organic Waste plate (deck location 3). After removing the cartridges at the seating station (deck location 2), the syringes are washed the specified number of cycles at the wash station (deck location 1) using 250 µL per wash cycle.
Note: The Priming Buffer (deck location 5) is also used for the Syringe Wash Buffer.
Always perform the Prime step at the beginning of a run. In the default protocol, this step is selected.
Table Setting the Prime step parameters
 
Equilibrate step
The Equilibrate step exchanges the Priming Buffer with Equilibration Buffer within the resin bed. Equilibrating the cartridge bed with a buffer that establishes optimal chemical conditions for sample binding is critical for the Load Samples step.
The Equilibrate step aspirates the specified volume of Equilibration Buffer (deck location 8) into the syringes, mounts the cartridges (deck location 2) onto the syringes, and then dispenses the buffer at the specified flow rate through the cartridges into the Organic Waste plate (deck location 3). After removing the cartridges at the seating station (deck location 2), the syringes are washed the specified number of cycles at the wash station (deck location 1) using 250 µL per wash cycle.
Note: The Equilibration Buffer (deck location 8) is also used for the Cartridge Wash Buffer.
Always perform the Equilibrate step after cartridge priming to establish optimum sample loading conditions. In the default protocol, this step is selected.
Table Setting the Equilibrate step parameters
 
Load Samples step
The Load Samples step aspirates the samples (deck location 4) into the syringes, and then performs an external syringe probe wash at the wash station to remove any sample remaining on the outside of the probes before mounting the cartridges at the seating station (deck location 2). The sample is dispensed through cartridges at the specified flow rate into the Flow Through Collection plate (deck location 7) if the Collect Flow Through step is selected, or into the Organic Waste plate (deck location 3). During an external cartridge tip wash, the cartridge tips dip into the wash station chimneys (deck location 1) to remove any sample on the outside of the cartridge. After removing the cartridges at the seating station (deck location 2), the syringes are washed the specified number of cycles at the wash station (deck location 1) using 250 µL per wash cycle.
The Phosphopeptide Enrichment protocol permits up to 1000 µL of sample to be dispensed through the cartridges. However, each syringe has a maximum capacity of 250 µL. For sample volumes >250 µL, the protocol will iteratively load equal volumes onto the cartridges during the Load Samples step.
For example, to load 280 µL of sample onto cartridges from U-bottom plates, the protocol will load the sample in two 140 µL loading loops. Similarly, to load 900 µL from a deep-well plate, the protocol will load the sample in four sample loading loops of 225 µL each.
The protocol uses the following formula to determine the number and volume of iterative load steps:
# of times to load = Total sample volume/250,
and the result is rounded up to nearest integer
Volume of each load = Sample volume/# of times to load
The protocol uses logic to limit the amount of sample that can be aspirated from specific labware types. For example, an error message will appear at the beginning of a run if the volume of the Load Samples step is 500 µL and the labware selection for the Samples plate at deck location 4 is a PCR plate. This logic minimizes the likelihood of overflowing samples into adjacent wells when the probes are immersed in the wells to aspirate sample. This logic also adjusts the well-bottom offset and facilitates sample volume-dependent head movement based on the volume in the Load Samples step.
For the maximum practical working volumes per well for specific labware at deck location 4, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
Always select the Load Samples step during a normal run. In the default protocol, this step is selected.
Table Setting the Load Samples step parameters
 
Collect Flow Through (from the Load Samples step)
Note: The Collect Flow Through step is an option only if the Load Samples step is conducted.
If you select the Collect Flow Through step, the sample flow-through from the Load Samples step is dispensed directly into the Flow Through Collection plate (deck location 7).
If the Load Samples step is conducted, but the Collect Flow Through step is not selected, the sample flow-through will be dispensed directly into the Organic Waste plate (deck location 3).
For the maximum practical working volumes per well for specific labware at deck location 7, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
Select the Collect Flow Through step if you want to collect the sample flow-through generated during the Load Samples step for downstream processing or for analysis.
Cup Wash step
The Cup Wash step removes the small volume of residual sample liquid that might remain above the upper cartridge membrane after sample loading. This step effectively rinses the cup portion of AssayMAP cartridges with buffer. The Cartridge Wash Buffer (deck location 8) is aspirated into the syringes, and then dispensed into the cartridge cups (deck location 2). This liquid, plus any residual liquid from samples, is aspirated from the cartridge cups into the syringes. The protocol ensures that no cartridges are stuck to the probes before dispensing the liquid into the Organic Waste plate (deck location 3), and then washing the syringes at the wash station (deck location 1).
Note: The Cartridge Wash Buffer (deck location 8) is also used for the Equilibration Buffer.
Select the Cup Wash step if there is residual sample liquid in the cartridge cups that might cause complications with analysis.
Table Setting the Cup Wash step parameters
 
Internal Cartridge Wash step
The Internal Cartridge Wash step aspirates Cartridge Wash Buffer (deck location 8) into the syringes. After mounting the cartridges at the seating station (deck location 2), the buffer is dispensed through the cartridges at the specified flow rate into the Flow Through Collection plate (deck location 7), if the Collect Flow Through step is selected, or into the Organic Waste plate (deck location 3). An external cartridge tip wash is performed at the wash station (deck location 1), where the cartridge tips dip into the chimneys to remove any sample on the outside of the cartridge. After removing the cartridges at the seating station (deck location 2), the syringes are washed the specified number of cycles at the wash station (deck location 1) using 250 µL per wash cycle.
Note: The Cartridge Wash Buffer (deck location 8) is also used for the Equilibration Buffer.
If the Load Samples step was selected, an additional 5 µL of Cartridge Wash Buffer is aspirated for the sample chase in addition to the specified volume for the Internal Cartridge Wash step. The sample chase uses the Load Samples flow rate to dispense buffer through the cartridges so that all sample loaded onto the cartridge has an opportunity to interact with the entire bed of the cartridge (1 bed volume = 5 µL). After the first 5 µL of buffer has been dispensed through the cartridge, the remaining buffer is dispensed through the cartridges at the Internal Cartridge Wash flow rate. If the Load Samples step is skipped, the sample chase step is also skipped.
Select the Internal Cartridge Wash step to minimize non-target elements from the sample matrix in the cartridge resin bed, which could possibly contaminate the eluate. In the default protocol, this step is selected.
Table Setting the Internal Cartridge Wash step parameters
 
Collect Flow Through step (from the Internal Cartridge Wash step)
Note: The Collect Flow Through step is an option only if the Internal Cartridge Wash step is conducted.
The Collect Flow Through step enables the wash flow-through from the Internal Cartridge Wash step to be directly eluted into the Flow Through Collection plate (deck location 7) during the Internal Cartridge Wash step. If the Internal Cartridge Wash step is conducted, but the Collect Flow Through step is not selected, the wash flow-through is dispensed directly into the Organic Waste plate (deck location 3).
For the maximum practical working volumes per well for specific labware at deck location 7, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
Select this step if you want to collect the flow-through generated during the Internal Cartridge Wash step and combine it with the flow-through generated during the Load Samples step for downstream processing or analysis. In the default protocol, this step is not selected.
Stringent Syringe Wash step
The Stringent Syringe Wash step comprehensively cleans the syringes with Syringe Wash Buffer (deck location 5) to reduce the potential for sample carryover. The step aspirates buffer into the syringes (deck location 5), moves the head moves above the Organic Waste plate (deck location 3), draws the buffer up to the maximum available volume of the syringe, and then moves the head down and dispenses the buffer into the Organic Waste plate. The head then moves to the wash station (deck location 1) and washes the syringes with 250 µL of liquid per wash cycle.
Note: The Syringe Wash Buffer (deck location 5) is also used for the Priming Buffer.
Select the Stringent Syringe Wash step when rigorous cleaning of the syringes is necessary to prevent carryover. In the default protocol, this step is selected.
Table Setting the Stringent Syringe Wash step parameters
 
Elute step
The Elute step uses Elution Buffer to elute bound analytes from AssayMAP cartridges. The Elute step aspirates the specified volume of Elution Buffer into syringes (deck location 6), mounts the cartridges at the seating station (deck location 2), and then dispenses the Elution Buffer through the cartridges at the specified flow rate into the Eluate Collection plate (deck location 9). An external cartridge tip wash is performed at the wash station (deck location 1), where the cartridge tips dip into the chimneys to remove any eluate on the outside of the cartridge. After removing the cartridges at the seating station (deck location 2), the head moves back to deck location 9 where eluates are mixed in the Eluate Collection plate. (If the total volume in the Eluate Collection plate is < 15 µL, the eluates will not be mixed). The head then moves to the wash station (deck location 1) and washes the syringes for the specified number of wash cycles using 250 µL per wash cycle.
Using sufficiently strong elution conditions and the default flow rate (5 µL/min), phosphopeptides can be fully eluted from cartridges in as little as 10 µL (approximately 2 cartridge column volumes).
The maximum volume of eluate that can be collected in the Eluate Collection plate (deck location 9) is dependent on the maximum practical working volume of the labware defined for the Phosphopeptide Enrichment protocol. The total volume in the Eluate Collection plate is determined as follows:
Total volume = Net Elution Volume (Elute volume - Eluate Discard volume) + Existing Collection Volume
For the maximum practical working volumes per well for specific labware at deck location 9, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
Always select the Elute step to elute bound analytes from cartridges in a normal run. In the default protocol, this step is selected.
Table Setting the Elute step parameters
 
Eluate Discard step
Note: The Eluate Discard step is an option only if the Elute step is conducted.
The Eluate Discard step permits a specified volume of the first eluate from the AssayMAP cartridges to be discarded. The Eluate Discard step aspirates the specified volume of Elution Buffer (deck location 6) from the Elute step into the syringes, mounts the cartridges at the seating station (deck location 2), and dispenses the Elution Buffer through the cartridges at the specified Elute flow rate into the Organic Waste plate (deck location 3) or the Flow Through Collection plate (deck location 7), if the Add to Flow Through step is selected. The remaining Elution Buffer is dispensed through cartridges at the Elute flow rate into the Eluate Collection plate (deck location 9).
For example, if you select the Elute, Eluate Discard, and Add to Flow Through steps and set the Elute volume = 20 µL and the Eluate Discard volume = 3 µL, the first 3 µL of eluate from the cartridges will be discarded to the Flow Through Collection plate (deck location 7) and the next 17 µL of eluate will be collected in the Eluate Collection plate (deck location 9).
For some experimental situations, minimizing the total volume of sample eluates is critical to downstream processing. This may be the case when the concentration of organic solvent must be diluted to suitable levels for reversed-phase LC/MS analyses, but the total sample volume must stay within the allowable injection volume in order to inject the entire sample. In this case, it can be valuable to discard the first few microliters of eluate from an AssayMAP cartridge, which do not contain measurable amounts of analyte.
For AssayMAP cartridges, the first 2 to 5 µL of eluate contains small or no measurable amounts of analyte. The volume of the analyte-free eluate will vary depending on both the strength of the elution buffer and the degree of analyte saturation of the cartridges. This value should be determined empirically for each combination of sample, buffer, and cartridge type for a given experiment.
Select the Eluate Discard step if minimizing the volume of eluate is critical. In the default protocol, this step is not selected.
Table Setting the Eluate Discard step parameters
 
Add to Flow Through step
Note: The Add to Flow Through step is an option only if both the Elute and Eluate Discard steps have been selected.
If selected, the Add to Flow Through step dispenses the Eluate Discard volume into the Flow Through Collection plate (deck location 7). If the Add to Flow Through step is not selected, the Eluate Discard is dispensed into the Organic Waste plate (deck location 3).
For the maximum practical working volumes per well for specific labware at deck location 7, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
Select the Add to Flow Through step if both the Elute and Eluate Discard steps have been selected and the Eluate Discard volume is to be added to the Flow Through Collection plate at deck location 7. In the default protocol, this step is not selected.
Existing Collection Volume step
This step enables users to specify an amount of liquid that is already contained in the wells of the Eluate Collection plate (deck location 9). The protocol uses the Existing Collection Volume and the net volume from the Elute step (Elute Volume - Eluate Discard Volume) to adjust the well-bottom offset for sample elution, calculate the eluate mixing volume, and dynamically move the head in and out of the wells during elution and eluate mixing in a volume-dependent manner.
For the maximum practical working volumes per well for specific labware at deck location 9, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
Select the Existing Collection Volume step if the Eluate Collection plate contains some volume of liquid useful for immediately diluting the eluates, for immediately adjusting the pH of the eluates, or to aid in the recovery of small volumes of eluates from the AssayMAP cartridges.
Table Setting the Existing Collection Volume step parameters
 
Final Syringe Wash step
The Final Syringe Wash step uses the wash station (deck location 1) to flush any potential contaminants from the syringes in the head. During each Final Syringe Wash cycle, the head aspirates 250 µL into the syringes from the wash station chimneys, and then moves by a fixed offset between the chimneys to dispense the syringe contents to waste.
Select the Final Syringe Wash step after each run to clean the syringes thoroughly before running a different assay. In the default protocol, this step is selected.
Table Setting the Final Syringe Wash parameter
 
The amount of carryover in the syringe (expressed as dilution of the starting sample) depends on the total number of syringe wash cycles. Three cycles give a dilution of up to 103. In cases where carryover is a major concern, increasing the number of iterations also increases the dilution factor, but with a cost of increased assay time and reduced syringe lifetime.
 
Return to top
Automation movements during the protocol
This section describes the basic automation movements of the AssayMAP Bravo Platform during the Phosphopeptide Enrichment protocol when the protocol defaults are used. Changing the selections or deselecting options will alter the movements. The default protocol takes approximately 50 minutes to complete.
 
Mounts one column of Fe(III)-NTA or TiO2 cartridges from column A of the 96AM Cartridge Rack and Receiving Plate.
Parks the cartridges in column A of the 96AM Cartridge & Tip Seating Station.
 
Return to top