|
• Prepare the Source Plate and Destination Plate with their intended reagents. Ensure the Source Plate has the required volume overage for the labware type. For details, see the Labware requirements table.The following table lists the labware options for Source Plate and Destination Plate at Bravo deck locations 7 and 8.
Agilent 201280-100 Agilent 201282-100 Agilent 201254-100 *For dimensionally equivalent alternatives, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
–
– Click Load to display all the settings associated with the selected method.
– To use a saved method exactly as written, skip to Running the Reagent Transfer utility. To modify the method, proceed to step 2 to make changes.
2 Under Application Settings area, specify the following:
If the Use Pipette Tips check box is selected, specify whether to transfer pipette tips using the Reagent Transfer utility.
• To transfer the pipette tips from a source tip box at deck location 6 to the empty 96AM Cartridge & Tip Seating Station at deck location 2, select the check box.
• To use pipette tips that are already present in the 96AM Cartridge & Tip Seating Station at deck location 2, clear the check box.For example, if you do a run and transfer the pipette tips from deck location 2 to the 96AM Cartridge & Tip Seating Station, and then want to do additional runs with the same setting, the pipette tips can be reused. If the Transfer Pipette Tips from Deck Location 6 check box is selected, specify the range of full columns of pipette tips present in the source tip box (deck location 6) at the beginning of the run.
When using pipette tips, this range can differ from the range specified for the Columns of Samples to Transfer value, but it must contain the same number of full columns. If not, and the software displays an error message, and the protocol automatically aborts.When using bare probes, use the default range of 1 to 12. The AssayMAP Bravo head moves as if it is transferring liquid from all 96 wells. The wells that contain liquid dictate which wells are transferred. where,
x is the final volume in the destination wells,
maximum Syringe Capacity = 250µL,
maximum Tip Capacity = 140µL
• If using pipette tips. This step occurs before the Initial Wash cycles.
• If using bare probes. This step occurs after the Initial Wash Cycles are completed.
• To specify forward pipetting mode. Use the default setting of 0 µL.
• To specify reverse pipetting mode. Use a value that is greater than 0 µL, and select the discard location for the extra liquid.
• Forward pipetting mode. The blowout occurs after the transfer has completed, and then again after mixing, if conducted.
• Reverse pipetting mode. The blowout occurs only after the mixing step, if conducted, because of the presence of the Reverse-Pipetting Overage volume.
• <Automatic> (default). Automatically assigns one of the following liquid classes, based on whether pipette tips are used and the volume being transferred. These are good general-purpose liquid classes for most reagents and buffers that are used for the AssayMAP system.
• Slow Flow (5 µL/sec). A slower flow rate is better for viscous solutions.
• Fast Flow (100 µL/sec). A faster flow rate is a good starting point for high organic solutions.If these flow rates do not provide the desired performance, you may enter a custom liquid class. To create a custom liquid class, you use the VWorks Liquid Library Editor. For details on how to open the VWorks Liquid Library Editor, see Creating a custom liquid class.In the Liquid Class box, type the liquid class name exactly as it appears in the VWorks Liquid Library Editor.
3 In the Labware Table, select that labware you are using at deck locations 7 and 8.
a
b Click Save to save the settings in a method file.
1
2
3 If you selected Use Pipette Tips, do one of the following:
• If you selected the Transfer Pipette Tips from Deck Location 6 check box, ensure that the:
– Empty 96AM Cartridge & Tip Seating Station is in place at deck location 2.
• If you cleared the Transfer Pipette Tips from Deck Location 6 check box, ensure that the 96AM Cartridge & Tip Seating Station at deck location 2 contains the specified range of pipette tips.
5 Click Run Protocol to start the protocol run.Figure. Error message examples
a Follow the on-screen instructions to click Continue to abort the run.
c After you have resolved the conflict, click Run Protocol to start the run.Figure. Warning message examples
• If the settings are correct, click Continue to start the run.
– Click Pause and Diagnose. The Scheduler Paused dialog box opens.
–
– Click Run Protocol to start the run.You can monitor the progress of the run in the Status box in the upper right corner of the form.To pause the run, click Pause. The task currently in progress finishes before the protocol pauses. The Scheduler Paused dialog box opens. For details, see Emergency stops and pauses.To troubleshoot errors, see the Error Recovery Guide and the Bravo Platform User Guide in the Literature Library page of the Protein Sample Prep Workbench.
1 In the Reagent Transfer v2.1 form, click Toggle Full Screen to change the display of the form so that the VWorks menubar and toolbar are visible, as the following figure shows.
2
3 Figure. VWorks Tools menu and form displayed with full screen off
Figure. VWorks Liquid Library Editor window
The following table describes the basic movements of the AssayMAP Bravo Platform when you run the default Reagent Transfer utility. Different Application Settings will alter the movements.
The Bravo 96AM Head moves above the wash station at deck location 1, and then purges the syringe contents to waste for the number of cycles set in the Syringe Drying Cycles setting. This step is necessary because the Bravo 96AM Head is regularly used to aspirate liquids directly into the syringes, and the presence of residual liquid can cause contamination or inaccurate pipetting behavior. The head picks up the pipette tips from the source tip box at deck location 6 and transfers them to the 96AM Cartridge & Tip Seating Station at deck location 2. The head picks up all the pipette tips from the 96AM Cartridge & Tip Seating Station at deck location 2.