AssayMAP Protein Sample Prep Workbench : Utilities: Normalization v2.0 User Guide : Before you start

Before you start
This topic lists the required hardware, software, labware, and solutions for running the Normalization protocol.
Hardware
Required hardware
 
 
The AssayMAP Bravo Platform is required for running the AssayMAP protocols. The following figure and table describe the platform components.
Figure. AssayMAP Bravo Platform components
 
 
Pump Module 2.0 and two carboys
Bravo Risers, 14.6 cm
Orbital Shaking Station with Control Unit
 
 
 
Optional equipment
You might need the following when preparing the samples and reagents:
Microplate centrifuge, such as the Agilent Microplate Centrifuge or equivalent
Microplate sealer, such as the PlateLoc Sealer or equivalent
 
Return to top
Software
 
 
The following table lists the minimum software requirements:
 
Agilent VWorks Automation Control
Agilent Protein Sample Prep Workbench
 
 
 
Return to top
Labware
The Normalization protocol requires the following labware:
 
*Using different pipette tips, such as filtered tips, will result in equipment damage. Ensure that you use the standard 250-µL pipette tip (Agilent 19477-002).
 
The following tables list the labware options, the corresponding deck locations, and the excess volume recommendations for each option.
Table Labware options for the Diluent reservoir and the corresponding excess volume recommendations
 
Agilent 201280-100
Agilent 201282-100
Table Labware options for Initial Sample plate (deck location 2) and Normalized plate (deck location 6)
 
The unusable volume (also known as dead volume) is the volume that cannot be reliably aspirated from the microplate well due to pipetting limitations that arise from differences in well-bottom geometry. The following table provides some general guidelines on how different microplate geometries can affect this value.
Table Guidelines on how microplate geometry can affect volume requirements
 
For dimensionally equivalent alternatives and other labware details, see the Labware Reference Guide in the Literature Library page of the Protein Sample Prep Workbench.
Return to top
Solutions
The Normalization utility requires the following solutions:
 
Return to top