![]() |
Figure. Normalization Method Setup Tool
The Method Setup Tool steps you through the process of defining a normalization method and creating a method file. The method file contains all the required information for the Normalization protocol on the AssayMAP Bravo Platform. The file is a record of the intended normalization process.
1
2
1 To clear any data from the Normalization Method Setup Tool, click Clear All (upper right corner).
2 Click Import Sample Concentrations. The Import Initial Sample Concentrations dialog box opens.
3 Click Import concentrations from an Excel file (.xls or .xlsx).
4
5 In the Excel file, select the 8-row by 12-column array of values to be imported, and then click OK in the Select Data dialog box.
1 Click Import Sample Concentrations. The Import Initial Sample Concentrations dialog box opens.
2 Click Import concentrations from a text file (.csv or .txt).
3
4 In the Excel Text Import Wizard that appears, follow the instructions to specify the type of delimited file.
5
6 In the imported worksheet, select the 8‑by‑12 array of values, and then click OK in the Select Data dialog box.
In the green box next to Step 2) Define Concentration Units, verify that the unit of measure for the sample concentration is correct. If necessary, type the correct units.
1 Click the Sample Plate box, and then select the labware from the list.
2 Notice the value that automatically appears in the Unusable Volume box. To calculate the usable volume, the calculator will subtract this value from the value in Step 4) Initial Sample Volume.The unusable volume (also known as dead volume) is the volume that cannot be reliably aspirated from the microplate well due to pipetting limitations that arise from differences in well-bottom geometry.The default value for Unusable Volume is a recommended setting that is based on extensive testing with the labware. If necessary, you may change the value, for example if the value is not appropriate for the sample liquid.Note: The Labware Reference Guide also presents labware-specific pipetting overage and maximum well capacity. You can find this guide in the Literature Library page of the Protein Sample Prep Workbench.
1 In the green (µL) box, type the initial sample volume.
2 Click Select Wells to Fill.
3 When the Initial Sample Volume dialog box opens:
a Select the array of wells to be filled in the Usable Sample Volumes plate grid of the worksheet.
b
4
2 Click Select Wells to Fill.
3 When the Target Sample Concentration dialog box opens:
a Select the array of wells to be filled in the Target Sample Concentrations plate grid of the worksheet.
b Click OK in the Target Sample Concentration dialog box to populate the selected wells of the plate grid.Note: If you want to clear all values from the Target Sample Concentrations plate grid, click Clear.
1 In the green (µL) box, type a value that you want to apply to multiple wells of the microplate.
2 Click Select Wells to Fill.
3 When the Target Final Volume dialog box appears:
a Select the array of wells to be filled in the Target Final Volumes plate grid of the worksheet.
b Click Calculate Volumes. An algorithm uses the information provided in the previous steps to calculate the sample and diluent volumes required to achieve the specified normalization targets. The calculated volumes (µL) are displayed in the following areas:
Any wells with red highlights are considered exceptions, for which the calculated sample or diluent volume values cannot be achieved because the volumes required are either larger than the volumes available in the Sample plate or smaller than practical for the AssayMAP Bravo Platform to transfer with high accuracy.You must click Check Exceptions in order to proceed, even if there are no red highlighted wells in the plate grid. The setup tool also performs other important conformity checks at this time.
1 Click Check Exceptions. The Manage Exceptions dialog box opens. The descriptions change in the Status and the Selecting Do Best Will areas depending on the conditions that caused the exception for the selected sample. The following figure shows an example.
1
3
The AssayMAP Bravo Platform will attempt to perform the normalization as defined, by pipetting less than 5 µL of sample.If you have exceptions, make sure you review the Expected Concentrations after Normalization values at the bottom of the worksheet before you create the method file.
1 Click the Normalized Plate green box, and then select the labware from the list.Note: This plate will be placed on a Bravo Plate Riser at deck location 6, which is where the final normalized samples will be at the end of the run.
2 Notice the value that automatically displays in the Maximum Volume (µL per well or channel) box.
3 Click the Diluent Reservoir green box, and then select the labware from the list.Note: This reservoir will be placed at deck location 4 without a Bravo Plate Riser. The reservoir will supply diluent for the normalization process.
4 Notice the corresponding value in the Maximum Volume (µL per well or channel) box. This value minus the dead volume associated with the selected plate type is the volume that the Bravo 96AM Head can access in the selected labware. The Bravo 96AM Head uses a single pipette tip mounted on probe A12 to aspirate the volume.
In this step you select the number of mix cycles to perform after all the samples and diluent are added in the Normalized plate. The Bravo 96AM Head presses on the full selection of used pipette tips from deck location 3, and then mixes all the wells at the same time.Click the mix cycles box, and then select the value from the list.
In general, a Normalization protocol run of 96 samples on the AssayMAP Bravo Platform requires approximately 110 minutes. During this time, the samples can evaporate, especially by the time the 96th sample is transferred. The optional Evaporation Correction Factor helps to compensate for the sample evaporation.The evaporation rate will vary depending on the sample solution. You should empirically determine the correction factor using test samples. See Assay development guidelines for a detailed explanation of when to use this factor and how to calculate an appropriate value to use.
1 In the Step 11) green box, type a value (%) for the correction factor, and then click Apply Correction Factor.The tool automatically adjusts the values in the Final Diluent Volumes and Final Sample Volumes plate grids. The button label changes to Undo Correction Factor.
1 Review the Diluent required for Diluent Reservoir area at the bottom of the worksheet. The Method Setup Tool calculates the total fill volume for the Diluent reservoir based on the calculated Final Diluent Volumes and the recommended excess volume for the selected labware type.
2 If a red highlight appears on the Total to fill Reservoir value, the diluent volume required is more than the selected labware can hold. In this case, select a different labware option for the Diluent Reservoir in Step 9).Note: If you are using a 12-column reservoir or an 8-row reservoir, the Total to fill Reservoir value represents the volume in the rightmost column or topmost row, respectively.
3 Make a note of the Total to fill Reservoir volume. You will use this value to prepare the diluent reservoir for the normalization run.
1 Click Create Method File.
2
3 To close the Normalization Method Setup Tool:
a Click File > Close or exit the Microsoft Excel software. A Microsoft Excel dialog box opens and asks if you want to save the changes.
b If you want the current values in the worksheet to remain each time you open the tool, click Save. Otherwise, click Don’t Save.