![]() |
The Method Setup Tool steps you through the process of defining a serial dilution method and creating a method file. The method file contains all the required information for the Serial Dilution protocol on the AssayMAP Bravo Platform. The file is a record of the intended serial dilution process.
1
If you change anything in step 1 or 2 after completing step 3 or 4, make sure that you click Clear Steps 3 and 4 Only and redo steps 3 and 4.
1 To clear any data from the Serial Dilution Method Setup Tool, click Clear All (upper right corner).
• <Automatic> (default). Automatically assigns one of the following liquid classes, based on the volume being transferred.
• Slow Flow (5 µL/sec). A slower flow rate is better for viscous solutions.
• Fast Flow (100 µL/sec). A faster flow rate may improve performance for high organic solutions.If these flow rates do not provide the desired performance, you may enter a custom liquid class. To create a custom liquid class, you use the VWorks Liquid Library Editor. For details on how to open the VWorks Liquid Library Editor, see To create a custom liquid class:.In the Liquid Class box, type the liquid class name exactly as it appears in the VWorks Liquid Library Editor. For options, see Labware. For the capacity limits and the volume overage recommended for each labware, see the Labware Reference Guide in the Literature Library page of the workbench. For options, see Labware. For the capacity limits and the volume overage recommended for each labware, see the Labware Reference Guide in the Literature Library page of the workbench.Be sure to choose a diluent reservoir that can hold the volume required for the serial dilution. The volume of diluent and the location where it should be placed will given in step 7 of the Method Setup Tool. The plate type chosen must have enough well capacity to hold the dilution indicated in step 7 plus the required well overage. For volume guidelines, see Preparing the diluent and samples.Agilent Technologies does not guarantee specific accuracy and precision results for the Serial Dilution v2.0 utility because accuracy and precision are too dependent on the composition and liquid properties of the solutions used.
1 In the Serial Dilution form, click Toggle Full Screen to change the display of the form so that the VWorks menubar and toolbar are visible, as the following figure shows.
2
3 Figure. VWorks Tools menu and form displayed with full screen off
Figure. VWorks Liquid Library Editor window
In the green boxes in the Final column, enter the desired concentration of each serial dilution step.The method setup tool automatically enters the dilution concentration for dilution 1 when you enter the Final concentration of dilution 1 (Step 1) Set dilution parameters). You enter all the other concentrations in the dilution series directly in each cell or by entering a formula in the cell. You can use the Excel fill series function with formulas to quickly generate a serial dilution.Note: If you enter zero as a concentration, an error message appears and says Negative or zero values cannot be plotted correctly on log charts. Only positive values can be interpreted on a logarithmic scale. To continue and use the zero, click OK. The graph will not include the zero volume but the method will work.
• If step 1 Optional volume to be added is >0, the Post Run column will show the concentration of the serial dilution steps after the serial dilution utility is run but before the optional volume is added. The concentrations entered in the Final column represent the concentration of the serial dilution after adding the optional volume.
• If step 1 Optional volume to be added is 0, the concentrations entered in the Final column will be the same as the those in the Post Run column.
If you change anything in step 1 or 2 after starting step 3 or 4, make sure that you click Clear Steps 3 and 4 Only and redo steps 3 and 4.
1 Click Select Starting Well (figure, item 1) and then click the corresponding cell in the Serial Dilution Plate map where the replicate 1 of dilution 1 will be located (figure, item 2).
2 When the Choose Starting Location dialog box opens, displaying a formula, click OK (figure, item 3).
Click Calculate Volumes. The tool uses the information provided in the previous steps to calculate the sample and diluent volumes that will be transferred into each well to achieve the specified serial dilution series.
If you change anything in step 1 or 2 at this point, click Clear Steps 3 and 4 Only and redo steps 3 and 4.
1 Click Create Serial Dilution Method.
2
2