AssayMAP Protein Sample Prep Workbench : Utilities: Serial Dilution v2.0 User Guide : Assay development guidelines

Assay development guidelines
This topic provides guidelines for adjusting the protocol, and notes about the application.
Note: For maximum well capacities and overage recommendations in the serial dilution plates, see the Labware Reference Guide in the Literature Library of the Protein Sample Prep Workbench.
Automation movements and stepwise guidelines
This section describes the basic automation movements of the AssayMAP Bravo Platform during the Serial Dilution run.
 
If 50 µL x Syringe/Tip Capacity,
then Mixing Volume = 0.75x
If x < 50 µL,
then Mixing Volume = 0.5x
where, x is the final volume in the destination wells, and
maximum 250 µL Tip Capacity = 140 µL