Before you start this procedure, make sure you are familiar with the workflow in Overview for setting sensor thresholds.
a In BenchCel Diagnostics, click the Controls tab. In the graphical display area click Stacker 1, and then click Move to sensors.
b
c In the Notch Sensors area, find the highest reading for the notch and lowest reading for the corner, and then record these readings. (In the example below, the highest notch reading is 7, and the lowest corner reading is 62.)
2 Find the highest notch reading (Notchhighest) among all the stacks. In the following example, the highest reading is 35 (stack 3).
3 Find the lowest corner reading (Cornerlowest) among all the stacks. In the following example, the lowest reading is 235 (stack 4).
5