Use this worksheet to record sensor readings and facilitate your threshold calculations. See Overview for setting sensor thresholds for the procedures to calculate the thresholds.Record the Plate presence reading for each stack.
Adjust the Intensity value so that the highest Plate presence reading is less than or equal to 175, and then record the Plate presence readings for each stack.
Highest reading among all the stacks (Punloaded): _________________After loading the microplates at each stack (Load All in BenchCel Diagnostics), record the Plate presence readings for each stack.
Lowest reading among all the stacks (Ploaded): _________________
In graphical display area on the Controls page, click Stacker, and then choose Move to Sensor. On the Jog/Teach tab, find the highest notch reading in the Notch Sensors area. Repeat this step for each stack.
Highest reading among all stacks (Notchhighest): ____________
Lowest reading among all stacks (Cornerlowest): ____________