About this topic
This topic explains how to use Bravo Diagnostics to configure a platepad to accept a lid in a delidding process.
Before you start
Configuring a platepad for delidding
To configure a platepad for delidding:
1
In Diagnostics, click the Profiles tab, and verify that the correct profile is initialized.
2
Click the Configuration tab.
3
In the Location is configured as list, select Accessory. The Accessories Wizard appears.
4
Choose the Location on which you want to place the microplate lid.
Note: Your selection in the Accessories Wizard overrides the selection in the Configuration tab.
Select the Lid hotel accessory to indicate that you want to use the Delidding and Relidding tasks.
Note: Because you can place only one lid on the designated platepad, Number of Slots is set to 1.
5
If you have not yet edited the teachpoint, click Yes. After the software initializes the accessory, edit the teachpoint.
6
In the Profiles tab, click Update this profile.