Bravo Platform User Guide : Diagnostics quick reference : Profiles tab quick reference

Profiles tab quick reference
About this topic
This topic provides a reference for the Profiles tab in Bravo Diagnostics.
Procedures
The Profiles tab enables you to:
 
Contents
The Profiles tab contains the following:
Profile Management area
The Profile Management area contains the following controls.
 
*Each profile can be used by multiple protocols. Deleting, renaming, or changing the parameters for a profile based on one protocol can invalidate other protocols that use the profile.
 
 
Initiates communication with the Bravo Platform using the selected profile.
Connection area
The Connection area contains the following controls.
 
Displays the device number of the Bravo device that is communicating with the computer.
Available only for earlier model devices. Specifies a serial communication connection between the device and the computer.
Miscellaneous area
The Miscellaneous area contains the following controls.
 
Specifies the z-axis distance or height (safety margin) above a microplate that the tips move to before moving to another quadrant of the microplate.
Sets the safe z-axis distance or height of the liquid-handling head.
A z-axis safe position of 0 is the highest position possible for the liquid-handling head. A z-axis safe position of 10 means the liquid-handling head is 10 mm below the highest point.
If you select the Always move to safe Z before each process option, the Bravo Platform moves to the z-axis safe height when moving between labware locations. Increasing this value may shorten run times by minimizing the z‑axis movement between tasks. Making this value too large causes the liquid-handling head to crash into labware when moving between plate locations.
Enables a warning to display when the Bravo Platform is initialized and before the w-axis is homed.
Moves the liquid-handling head to the specified z-axis safe position between processes.
If you select this option, the Bravo head moves to the specified z-axis safe height when moving between labware locations. Select this option if you have not specified the labware on the deck.
If you do not select this option, the Bravo Platform will automatically determine the safe z‑axis point based on the labware specified, thereby optimizing the processing time.
To prevent damage to the labware or the Bravo Platform components, do not select this option if you have not specified the labware.
Bravo Platform with gripper only. Enables the software to ignore the microplate sensor during a pick-and-place movement.
This is a Bravo SRT
Bravo SRT model only. Specifies that this Bravo device is the SRT model, which is shorter than the standard model.
Head Information area
The Head Information area contains the following controls.
 
CAUTION If this check box is cleared, the Bravo Platform will run with any liquid-handling head installed. Therefore, if the profile specifies a 96-channel head type, but a 384-channel head is installed, the head will crash.
Note: If this check box is cleared, the w-axis controls in the Jog/Teach tab will be disabled even when the device is initialized.
Modified Variables area
The Modified Variables area displays any changes to the profile settings since the last time the profile was saved.
Related information
 
Opening Diagnostics