This topic describes the task parameters for the Dispense to Waste task, which is available only for a location with a Tip Wash Station (also known as a MicroWash Reservoir).The Dispense to Waste task moves the pipettes by a horizontal offset and then dispenses used fluid in between the chimneys in the Tip Wash Station. The software calculates the horizontal offset automatically based on the labware definition for the Tip Wash Station.To run the Dispense to Waste task from Diagnostics, see Performing a task using Bravo Diagnostics. To use the task in a protocol, see the VWorks Automation Control User Guide.
The option to empty all the liquid from the head into the Tip Wash Station outside the chimneys. A JavaScript variable enables the value to be assigned later, for example through a VWorks form. Using a VWorks form, an operator could easily change the flow rate for a dispense step in increments from as low as 1 µL/min up to 1001 µL/min using the same liquid class.If the task is included in a VWorks macro, a JavaScript variable enables you to change the value for the task at the macro level. The absolute distance between the end of the pipette tips and the bottom of the Tip Wash Station during the Dispense to Waste task.
• At the same rate as the volume change. Calculate dynamic tip retraction (DTR) as follows:
• Faster than the volume change.
DTR > 1/A
• Slower than the volume change.
DTR < 1/A The option to touch the tip on an outer side of the adjacent northeast chimney or wall in the Tip Wash Station after performing the dispense.
Opening Diagnostics