This topic describes the task parameters for the AM Dispense task, which is available only for the Bravo 96AM Head.The AM Dispense task is designed for dispensing fluids through AssayMAP Bravo cartridges that are mounted on the Bravo 96AM Head. In addition to the parameters for the Dispense task, the AM Dispense task allows you to override the flow rate from liquid class.To run the AM Dispense task from Diagnostics, see Performing a task using Bravo Diagnostics. To use the task in a protocol, see the VWorks Automation Control User Guide.
A JavaScript variable enables the value to be assigned later. For example, using a VWorks form, an operator could easily change the flow rate for an aspirate step in increments from as low as 1 µL/min up to 2000 µL/min or more using the same liquid class.If the task is included in a VWorks macro, a JavaScript variable enables you to change the value for the task at the macro level. The distance between the tips of the probes, cartridges, or disposable tips and the well bottoms during the AM Dispense task.If the task is performed in the upper cup of unmounted cartridges, you can use a negative number for the parameter value. In this case, the well bottom is the top of the cartridge cup, as shown in the following figure.
• At the same rate as the volume change. Calculate dynamic tip retraction (DTR) as follows:
• Faster than the volume change.
DTR > 1/A
• Slower than the volume change.
DTR < 1/A
Opening Diagnostics