Bravo Platform User Guide : Diagnostics quick reference : AM Mix task parameters

AM Mix task parameters
About this topic
This topic describes the task parameters for the AM Mix task, which is available only for the Bravo 96AM Head.
The AM Mix task is designed for aspirating and dispensing fluids through AssayMAP Bravo cartridges that are mounted on the Bravo 96AM Head.
To run the AM Mix task from Diagnostics, see Performing a task using Bravo Diagnostics. To use the task in a protocol, see the VWorks Automation Control User Guide.
Parameter description
The AM Mix task has the following parameters.
 
 
 
A JavaScript variable enables the value to be assigned later. For example, using a VWorks form, an operator could easily change the flow rate for a mix step in increments from as low as 1 µL/min up to 2000 µL/min or more using the same liquid class.
If the task is included in a VWorks macro, a JavaScript variable enables you to change the value for the task at the macro level.
The rate at which the Bravo 96AM Head moves during the aspirate action. The software calculates the distance over which the tips will move without crashing.
At the same rate as the volume change. Calculate dynamic tip extension (DTE) as follows:
(Vaspirated * DTE) + Distancewell bottom
 
 
Related information
 
Opening Diagnostics