Labware MiniHub User Guide : Maintenance and troubleshooting : Troubleshooting error messages

Troubleshooting error messages
 
 
About this topic
The following tables list commonly encountered error messages, the causes of the errors, and ways to resolve the errors.
If you are still experiencing problems with the Labware MiniHub after trying the solutions, or if an error not on the list is displayed, contact Agilent Automation Solutions Technical Support.
For protocol-related errors, see the applicable automation control software user guide, such as the VWorks Automation Control User Guide.
Communication errors
The following error messages are caused by communication problems:
Check the following causes and try the corresponding solutions:
 
Motor control errors
The following error messages are typically caused by an obstacle in the path of the Labware MiniHub rotation.
Check the following causes and try the corresponding solutions:
 
An obstacle outside the Labware MiniHub is preventing it from rotating.
Remove any obstacles that are preventing the Labware MiniHub from rotating.
Try rotating the Labware MiniHub again.
The robot teachpoints might be set incorrectly, the Labware MiniHub has moved after the teachpoints were set, or the shelves are bent.
Check the robot teachpoints at the Labware MiniHub to ensure that the robot is placing the labware in the shelves correctly. See Setting robot teachpoints.
Related information
 
Labware MiniHub component names