This topic describes how to set the air pressure after connecting the air tubing to the Microplate Labeler.
1 Turn on the air supply to the Microplate Labeler.Depending on your lab configuration, the mechanism for turning on the air supply might be a pressure regulator located on the wall or on a compressed air cylinder, or a toggle switch on the air line. The air supply to the Microplate Labeler should be in the range of 0.55 MPa (80 psi) to 0.58 MPa (85 psi).
2 To adjust the internal air-pressure regulator on the Microplate Labeler:
Check the air pressure after using the Microplate Labeler for the first time to make sure the air pressure setting is correct.