Microplate Labeler User Guide : Installing the device : Connecting the device to the computer

Connecting the device to the computer
About this topic
This topic describes how to connect the computer to the Microplate Labeler using either of the following:
Using an Ethernet connection
You can use Ethernet to connect a Microplate Labeler to the computer using one of the following connections:
Direct connection. A communication cable connects a single Microplate Labeler directly to a single host computer, where no other robots or devices must interact or communicate.
Ethernet switch. An Ethernet switch connects the single cable from the computer to one or more cables that lead to one or more devices. If you are setting up a local area network (LAN) that has other devices on it, use an Ethernet switch.
The Agilent configured computer has two Ethernet ports:
Port for the add-on Ethernet card. Configured for a static address. You connect the Microplate Labeler to the port configured for a static address.
Port for the computer’s built-in Ethernet card. Configured for a dynamic address. You can use this port to connect to your company network.
To connect directly to the computer using Ethernet:
Connect one end of the Ethernet cable to the Ethernet port on the Microplate Labeler connection panel, and connect the other end of the cable to the Ethernet port of the computer.
Figure. Connection panel: Ethernet connection
 
 
To connect through an Ethernet switch:
1
Note: If you are adding the Microplate Labeler to an existing LAN, this step should already be done.
a
b
2
Connect the Microplate Labeler to the switch as follows:
a
b
Connecting additional devices
You can connect additional Microplate Labelers to the network through the switch.
Each Microplate Labeler requires an additional profile.
 
*Before connecting multiple Microplate Labeler devices on the same network, contact Agilent Automation Solutions Technical Support for guidance on setting the IP addresses.
 
About configuring the computer’s network card (Ethernet only)
The Agilent configured computer is already set up to communicate with the Microplate Labeler. No change to the network card IP address is required.
If you are using a computer other than an Agilent configured computer, make sure the value of the network card IP address and subnet mask are as follows:
If your computer will be connected to your LAN, make sure the computer has a second network card. The second network card can have a dynamic IP address.
Using a serial connection
If you are connecting a single Microplate Labeler to a computer, you can use a serial connection.
 
*Using serial connections to connect more than one Microplate Labeler to the same computer can cause problems with the barcode label formats.
 
To use a serial connection:
1
 
*The serial cable is a straight-through, pin-to-pin cable. Make sure you are not using a serial null-modem cable, which may look identical but will prevent proper communication with the Microplate Labeler.
 
2
Figure. Connection panel: serial connection
 
To install the automation control software, see the software release notes, for example, the VWorks Automation Control Software Release Notes.
Related topics
 
Agilent Automation Solutions Technical Support