This topic describes commonly-encountered error messages for the Microplate Labeler and how to resolve them.If your problem is not listed, or you are not able to resolve the problem, contact Agilent Automation Solutions Technical Support. See Reporting problems.
In Microplate Labeler Diagnostics, check the air pressure reading on the I/O Functions tab. In Microplate Labeler Diagnostics, check the air pressure reading on the I/O Functions tab.
Make sure that you are using the ribbon and labels provided in the Microplate Labeler Media Kit.See Loading labels and ribbon for installation instructions. In Microplate Labeler Diagnostics, check the air pressure reading on the I/O Functions tab.
Verify that the computer was connected correctly to the serial port on the Microplate Labeler connection panel (at the back of the applicator).
1 In Microplate Labeler Diagnostics, click the Profiles tab. At the Microplate Labeler connection panel (on the back of the applicator), press the power switch to the off (o) position, wait a full 5 seconds, and then press the power switch to the on (|) position.
Open Microplate Labeler Diagnostics and check label format. Make sure that you are using the ribbon and labels provided in the Microplate Labeler Media Kit.See Loading labels and ribbon for installation instructions. At the Microplate Labeler connection panel (on the back of the applicator), press the power switch to the off (o) position, wait a full 5 seconds, and then press the power switch to the on (|) position.