Vertical Pipetting Station User Guide : Installing Vertical Pipetting Station : About mounting the Vertical Pipetting Station

About mounting the Vertical Pipetting Station
About this topic
If the Vertical Pipetting Station is part of an Agilent Technologies lab automation system, such as the BioCel System, the unit is already mounted.
This topics describes the requirements for mounting the Vertical Pipetting Station on a lab bench or table or as part of a third-party lab automation system.
Mounting requirements
 
WARNING The Vertical Pipetting Station weighs 37.195 kg (82 lb). Do not attempt to move the Vertical Pipetting Station without assistance. Use proper lifting techniques when lifting the Vertical Pipetting Station.
Ensure the destination lab bench meets the requirements listed in General bench requirements.
If the Vertical Pipetting Station is operated as a standalone device, you are not required to fasten the Vertical Pipetting Station to the lab bench or table. Place the Vertical Pipetting Station next to the computer on the lab bench or table.
If the Vertical Pipetting Station is integrated as part of a third-party lab automation system, the Vertical Pipetting Station must be in a fixed position so that the other devices in the system can train teachpoints specific to the location of the Vertical Pipetting Station shelves. For mounting instructions, contact Automation Solutions Technical Support.