|
This topic provides instructions on how to connect the Vertical Pipetting Station to a grounded power source and to its controlling computer.Before you plug in the Vertical Pipetting Station, make sure the laboratory set up requirements have been met. See Laboratory requirements.
CAUTION Operating the Vertical Pipetting Station at the wrong voltage might seriously damage the device.Figure. Vertical Pipetting Station rear panel
To connect the Vertical Pipetting Station:
1 Connect the Vertical Pipetting Station power cable by plugging in one end of the cable at the AC power connector on the rear panel as the figure shows. Plug in the other end of the cable at an appropriately grounded electrical socket.
2 Connect the Vertical Pipetting Station serial communications cable by plugging in one end of the cable at the serial communications port on the Vertical Pipetting Station rear panel as the figure shows. Plug in the other end of the cable to the computer that controls the Vertical Pipetting Station.
3 If you are using the Automation Solutions Pump Module, see Setting up a Pump Module for connection details.
How to set up the Vertical Pipetting Station