Vertical Pipetting Station User Guide : Maintaining Vertical Pipetting Station : Recovering from a power outage

Recovering from a power outage
About this topic
A loss of power can be the result of either inadvertently turning off the power switch or a power failure. Regardless of the cause, a loss of power disrupts the communication between the Vertical Pipetting Station and the automation control software. You cannot resume a protocol that was running at the time of the power outage. To recover from the power outage safely, perform the following procedure.
 
Recovering from a power outage
To recover from a power outage:
1
On the Vertical Pipetting Station rear panel, ensure the power switch is set to the off (o) position and the air switch is set to the off (down) position.
 
2
If necessary, you can manually move a shelf to the out position.
3
 
For example, click Abort if an error message appears with options to Retry, Ignore, or Abort.
Alternatively, click Pause on the toolbar, and then click Abort process.
4
5
Open Vertical Pipetting Station Diagnostics and initialize the correct profile. Ignore the low air-pressure warning. Wait until the pipette head moves upwards in the z‑axis to the top of its travel.
6
The shelves should move to the outer positions.
7
8
Related topics
 
Opening Vertical Pipetting Station Diagnostics