Vertical Pipetting Station User Guide : Maintaining Vertical Pipetting Station : Troubleshooting hardware problems

Troubleshooting hardware problems
About this topic
This topic lists some potential hardware problems, the possible causes, and ways to resolve the problems.
 
 
Hardware problems
Locate your problem in the table and try the solution. If the problem persists after you try the solutions, contact Automation Solutions Technical Support.
 
The Vertical Pipetting Station does not turn on.
Ensure the power cord is in good condition and is connected to the Vertical Pipetting Station and the power source.
The Vertical Pipetting Station fuse is bad.
Check the air tubing and the connections on the Vertical Pipetting Station rear panel and at the source for leaks.
The Vertical Pipetting Station does not dispense accurately on a particular shelf.
Contact Automation Solutions Technical Support.
The tubing is kinked or the Weigh Shelf might require recalibration.
During initialization, a z‑position timeout occurs.
The z‑motor brake is sticking.
Contact Automation Solutions Technical Support.
 
 