VWorks User Guide : Setting parameters for microplate storage tasks : Storage Incubate

Storage Incubate
Description
The Storage Incubate task () moves a defined set of labware into a storage device, leaves it there for a specified time period, and then moves it out of the storage device.
Note:  
To incubate labware on a platepad or in a Vertical Pipetting Station, use the Incubate task.
To incubate labware on the Bravo Platform, use the Reserve Location task.
 
Task parameters
After adding the Storage Incubate task at the desired point in the protocol, set the following parameters in the Task Parameters area:
 
 
Double-click a location in the Available locations area. The selected location appears in the Assigned locations area.
Double-click a location group in the Available locations area. The selected location group appears in the Assigned locations area.
Example: Incubate destination microplates in the Labware MiniHub
Goal
Aspirate contents from two source microplates and dispense into a destination microplate. Incubate the destination microplate in the Labware MiniHub for 5 minutes and 10 seconds before reading the microplates.
Implementation
In the protocol, add a liquid-handling subprocess to specify the tasks for moving contents from the two microplates into the destination microplates. Add a Storage Incubate task after the liquid-handling subprocess and before the microplate-reading task as shown in the following example.
A
In the Storage Incubate Task Parameters area, set the Length of incubation at 5 minutes 10 seconds as shown in the following example. The example assumes that cassette 2 is available for use.
A
Related information