Emergency stops and pauses
Stopping in an emergency
To stop in an emergency:
Press the red button on the emergency-stop pendant. The Bravo head stops immediately.
The robot disable is active error message opens.
To restore the Bravo device after an emergency stop:
1 At the emergency-stop pendant, turn the red button clockwise. The spring-loaded button pops up.
2 If applicable. Remove any object that is interrupting the Light Curtain.
3 In the AssayMAP Bravo Error dialog box, click Retry.
In most cases, the AssayMAP Bravo Platform will be able to resume the run where it left off.
Pausing a run
To pause and then continue a run:
1 In the workbench application or utility control panel, click Pause.
The task currently in progress finishes before the protocol pauses. The Scheduler Paused dialog box opens.
2 While the Bravo device is idle, make the necessary changes to your run, for example, adjusting a labware position or volume.
3 To resume the run, click Continue in the Scheduler Paused dialog box.
To cancel the run, click Abort process in the Scheduler Paused dialog box.