Mix task parameters
About this topic
The Mix task aspirates and dispenses liquid multiple times to mix it. You can specify different well-bottom distances for the aspirate and dispense actions. This topic describes the task parameters.
Parameter descriptions
The Mix task parameters include the following.
 
 
 
Parameter
Description
Volume
(0–200 µL)
The volume of liquid to be mixed in each well.
Pre-aspirate volume
(0–200 µL)
The volume of air to be drawn before the pipette tips enter the liquid.
Blowout volume
(0–200 µL)
Specifies the volume of air to dispense after the main volume has been dispensed while the tips are still in the wells.
Typically, the blowout volume is the same as the pre-aspirate volume.
Liquid class
The pipetting speed and accuracy.
*To ensure consistent pipetting, always select a liquid class for liquid-handling tasks.
Mix cycles ((0–100)
The number of times to repeat the aspirate-and-dispense cycle.
Aspirate distance
(0–100 mm)
The distance between the end of the pipette tips and the well botttoms during the aspirate action.
*The labware definition must be accurate and the teachpoint must be precise in order for the system to position the tips at the correct distance from the well bottom.
Dispense at different distance
The option to dispense at a pipette tip height that is different than the aspirate distance.
Select the check box to enter a value for the dispense distance.
Dispense distance
(0–100 mm)
The distance between the end of the pipette tips and the well bottoms during the dispense action.
Dynamic tip extension
(0–20 mm/µL)
The rate at which the pipette head moves during the Aspirate task. The software calculates the distance over which the tips will move without crashing.
Use dynamic tip extension to prevent spills as the pipette tips displace the liquid.
To move the tips:
At the same rate as the volume change. Calculate dynamic tip extension (DTE) as follows:
DTE = (well depth)/(well vol) = 1/A, where A is the cross-sectional area of a well with straight walls
Faster than the volume change.
DTE > 1/A
Slower than the volume change.
DTE < 1/A
The starting and ending positions can be calculated as follows:
(Vaspirated * DTE) + Distancewell bottom
Note: Instead of a negative aspirated volume, the software automatically moves downward toward the well bottom with each aspirate action.
Pipette technique
The pipette location offset you want to use for the Dispense task.
The list of pipette techniques are defined in the Pipette Technique Editor.
Perform tip touch
The option to touch the pipette tip on one or more sides of the well.
Which sides to use for tip touch
The side or sides of the well to use during tip touch: North, South, East, West, North/South, West/East, West/East/South/North.
Tip touch retract distance
(–20 to 50 mm)
The vertical distance for the pipette tips to rise before touching the sides of the wells.
Tip touch horizontal offset
(–5 to 5 mm)
The horizontal distance the tips move. The value is based on the well diameter specified by the labware definition.
For example, if you set a value of:
0, the tips move a horizontal distance equal to the well radius
> 0, the tips attempt to move past the well radius, which results in a more forceful tip touch
< 0, the tips move a distance less than the radius of the well, resulting in a lighter tip touch
 
 
Related information
 
For information about…
See…
Opening Diagnostics
Editing the labware classes, liquid library, or pipetting techniques
VWorks Automation Control Setup Guide
Performing a diagnostic task
Reporting a problem