Parameter | Description |
---|---|
Reservoir mode | The action of the task: • Fill • Empty |
Pump speed | The speed, in percent of maximum, at which to pump the reagent. |
Pump on time | The duration of time, in seconds, to run the pump: If the Use weigh station/shelf option is selected, this value specifies the maximum time (seconds) to run the pump. If the pump runs for the full duration without reaching the weigh station/shelf threshold, an error occurs. |
Use weigh station/shelf | The option to use the Weigh Station or Weigh Shelf. |
Weigh station/shelf action threshold | The minimum fluid weight, in percent of the full weight that was calibrated on the Weigh Station or Weigh Shelf. For example, you can set the minimum threshold at 45% so that when the fluid reaches 45% of the full weight, fluid starts to pump into the reservoir. |
Weigh station stop action threshold | The maximum fluid weight, in percent of the full weight that was calibrated on the Weigh Station or Weigh Shelf. For example, you can set the stop threshold at 60% so that when the fluid reaches 60% of the full weight, fluid starts to drain or pump out of the reservoir. |