Parameter | Description |
---|---|
Volume (0–250 µL) | The volume of liquid to be drawn into each probe, cartridge, or tip. |
Pre-aspirate volume (0–250 µL) | The volume of air to be drawn before the probes, cartridges, or tips enter the liquid. |
Post-aspirate volume (0–250 µL) | The volume of air to be drawn after the liquid is drawn. |
Liquid class | The pipetting speed and accuracy. ![]() |
Override aspirate flow rate from liquid class | The option to override the aspirate velocity in the specified liquid class. Selecting this option enables you to specify a value for the aspirate flow rate without changing the liquid class. |
Aspirate flow rate (1e-05 – 500 µL/s) | The numerical value or the JavaScript variable that will override the aspirate velocity setting in the liquid class. A JavaScript variable enables the value to be assigned later. For example, using a VWorks form, an operator could easily change the flow rate for an aspirate step in increments from as low as 1 µL/min up to 2000 µL/min or more using the same liquid class. If the task is included in a VWorks macro, a JavaScript variable enables you to change the value for the task at the macro level. ![]() |
Distance from well bottom (-20–100 mm) | The distance between the tips of the probes, cartridges, or disposable tips and the well bottoms during the AM Aspirate task. If you specify dynamic tip extension, this is the distance at the end of the AM Aspirate task. Use a positive number for tasks that are performed with mounted cartridges, mounted tips, or bare probes at labware other than a cartridge rack. If the task is performed in the upper cup of unmounted cartridges, you can use a negative number for the parameter value. In this case, the well bottom is the top of the cartridge cup, as shown in the following figure. ![]() For example, a negative number enables a cup wash step while the cartridges are in the cartridge rack before they are mounted on the Bravo 96AM Head. ![]() ![]() |
Dynamic tip extension (0–20 mm/µL) | The rate at which the Bravo 96AM Head moves during the AM Aspirate task. The software calculates the distance over which the tips will move without crashing. Use dynamic tip extension to prevent spills as the pipette tips displace the liquid. To move the tips: • At the same rate as the volume change. Calculate dynamic tip extension (DTE) as follows: DTE = (well depth)/(well vol) = 1/A, where A is the cross-sectional area of a well with straight walls • Faster than the volume change. DTE > 1/A • Slower than the volume change. DTE < 1/A The starting and ending positions can be calculated as follows: (Vaspirated * DTE) + Distancewell bottom Note: Instead of a negative aspirated volume, the software automatically moves downward toward the well bottom with each aspirate action. |
Pipette technique | The pipette location offset you want to use for the AM Aspirate task. The list of pipette techniques are defined in the Pipette Technique Editor. |
Perform tip touch | The option to touch the pipette tip on one or more sides of the well. |
Which sides to use for tip touch | The side or sides of the well to use during tip touch: North, South, East, West, North/South, West/East, West/East/South/North. |
Tip touch retract distance (–20 to 50 mm) | The vertical distance for the pipette tips to rise before touching the sides of the wells. |
Tip touch horizontal offset (–5 to 5 mm) | The horizontal distance the tips move. The value is based on the well diameter specified by the labware definition. For example, if you set a value of: • 0, the tips move a horizontal distance equal to the well radius • > 0, the tips attempt to move past the well radius, which results in a more forceful tip touch • < 0, the tips move a distance less than the radius of the well, resulting in a lighter tip touch |
For information about… | See… |
---|---|
Opening Diagnostics | |
Editing the labware classes, liquid library, or pipetting techniques | VWorks Automation Control Setup Guide |
Performing a diagnostic task | |
Reporting a problem |