AM Mix task parameters
About this topic
This topic describes the task parameters for the AM Mix task, which is available only for the Bravo 96AM Head.
The AM Mix task is designed for aspirating and dispensing fluids through AssayMAP Bravo cartridges that are mounted on the Bravo 96AM Head.
To run the AM Mix task from Diagnostics, see Performing a task using Bravo Diagnostics. To use the task in a protocol, see the VWorks Automation Control User Guide.
Parameter description
The AM Mix task has the following parameters.
 
 
 
Parameter
Description
Volume (0–250 µL)
The volume of liquid to be mixed in each well.
Pre-aspirate volume (0–250 µL)
The volume of air to be drawn before the pipette tips enter the liquid.
Blowout volume (0–250 µL)
The volume of air to dispense after the main volume has been dispensed while the tips are still in the wells.
Typically, the blowout volume is the same as the pre-aspirate volume.
Liquid class
The pipetting velocity and accuracy.
*To ensure consistent pipetting, always select a liquid class for liquid-handling tasks.
Override flow rates from liquid class
The option to override the aspirate and dispense velocity in the selected liquid class. Selecting this option enables you to specify a value for the flow rate without changing the liquid class.
Aspirate/dispense flow rate (1e-05–500 µL/s)
The numerical value or the JavaScript variable that will override both the aspirate and dispense velocity in the selected liquid class.
A JavaScript variable enables the value to be assigned later. For example, using a VWorks form, an operator could easily change the flow rate for a mix step in increments from as low as 1 µL/min up to 2000 µL/min or more using the same liquid class.
If the task is included in a VWorks macro, a JavaScript variable enables you to change the value for the task at the macro level.
*The software requires that the flow rate value be in microliters per second (µL/s) at run time. If you want an operator to enter the value in microliters per minute (µL/min), you can use scripting to convert the values for the software to use.
Mix cycles (0–100)
The number of times to repeat the aspirate-and-dispense cycle.
Dynamic tip extension
(0–20 mm/µL)
The rate at which the Bravo 96AM Head moves during the aspirate action. The software calculates the distance over which the tips will move without crashing.
Use dynamic tip extension to prevent spills as the pipette tips displace the liquid.
To move the tips:
At the same rate as the volume change. Calculate dynamic tip extension (DTE) as follows:
DTE = (well depth)/(well vol) = 1/A, where A is the cross-sectional area of a well with straight walls
Faster than the volume change.
DTE > 1/A
Slower than the volume change.
DTE < 1/A
The starting and ending positions can be calculated as follows:
(Vaspirated * DTE) + Distancewell bottom
Note: Instead of a negative aspirated volume, the software automatically moves downward toward the well bottom with each aspirate action.
Pipette technique
The pipette location offset you want to use for the Mix task.
The list of pipette techniques are defined in the Pipette Technique Editor.
Aspirate distance
(0–100 mm)
The distance between the end of the pipette tips and the well botttoms during the aspirate action.
*The labware definition must be accurate and the teachpoint must be precise in order for the system to position the tips at the correct distance from the well bottom.
Dispense at different distance
The option to dispense at a pipette tip height that is different than the aspirate distance.
Select the check box to enter a value for the dispense distance.
Dispense distance
(0–100 mm)
The distance between the tips of the probes, cartridges, or disposable tips and the well bottoms during the dispense action.
Perform tip touch
The option to touch the tips on one or more sides of the well.
Which sides to use for tip touch
The side or sides of the well to use during tip touch: North, South, East, West, North/South, West/East, West/East/South/North.
Tip touch retract distance
(–20 to 50 mm)
The vertical distance for the tips to rise before touching the sides of the wells.
Tip touch horizontal offset
(–5 to 5 mm)
The horizontal distance the tips move. The value is based on the well diameter specified by the labware definition.
For example, if you set a value of:
0, the tips move a horizontal distance equal to the well radius
> 0, the tips attempt to move past the well radius, which results in a more forceful tip touch
< 0, the tips move a distance less than the radius of the well, resulting in a lighter tip touch
 
 
Related information
 
For information about…
See…
Opening Diagnostics
Editing the labware classes, liquid library, or pipetting techniques
VWorks Automation Control Setup Guide
Performing a diagnostic task
Reporting a problem