Emergency stops
When to use the emergency stop procedure
Use this procedure to:
Abort a head movement immediately
Abort a run in an emergency situation
If the emergency-stop button is pressed while the head is aspirating or dispensing, the pipetting accuracy might be impaired. To pause and then continue a run, use the Pause button in the VWorks software.
Procedures
To stop the head motion in an emergency:
1 Press the red button on the emergency-stop pendant. The Bravo motion stops.
 
To recover from an emergency stop:
2 Turn the red button on the emergency-stop pendant clockwise to release it.
 
3 In the VWorks message box, click one of the action buttons to re-enable the motors in the head.
 
If you want to…
Then click…
Abort the command or task
Abort
Attempt re-execution of the command or task
Retry
Ignore the current command or task and proceed to the next command or task
Ignore
Related information
 
For information about…
See…
Pausing and continuing a run
VWorks Automation Control User Guide
Safety hazards
Automation Solutions Products General Safety Guide
Safety and installation guide for your model Bravo Platform, for example, G5562A, G5563A Bravo Platform Safety and Installation Guide