Agilent recommends that you schedule an annual preventative maintenance. Preventive maintenance is an important part of keeping the Bravo Platform running smoothly and error free. To schedule the annual preventive maintenance service, contact Agilent Technical Support.Clean the Bravo Platform weekly or as needed. Use standard laboratory wipes and a mild detergent or isopropyl alcohol to clean the exterior painted white surfaces and the metal surfaces of dust, grime, chemical deposits, and other debris.
Inspect the Bravo Platform for wear. To resolve these issues, contact Agilent Technical Support. To resolve these issues, contact Agilent Technical Support. Verify teachpoint accuracy for each deck location. Reset the teachpoints as required. Pump Module. If applicable, inspect the tubing. Change out the tubing as necessary.For instructions on routing the tubing, see the Pump Module User Guide. Weigh Station. If applicable, verify that the configured empty and full weights are still valid.
Cleaning the Bravo Platform