|
This topic describes the most common error messages that might be encountered with the Bravo Platform and provides some possible solutions.
The Verify that it is safe to home the W-axis message appears during initialization. The Prompt user to home W‑axis on first initialization option is selected in the profile. The There appears to be a plate present error message appears during initialization. The Ignore plate sensor during pick and place option is not selected in the profile. If the gripper is not gripping labware, click Ignore and Continue to continue the homing process.
• Turn off the Bravo Platform off, and then turn it back on.
• Initialize the Bravo Platform. The Bravo Platform does not appear in the Discovered BioNet Devices dialog box. (Diagnostics > Profiles > Find available device). Bravo Platform is not communicating with VWorks software.
1
2 Turn off the Bravo Platform, and then turn it back on.
3 Click Refresh again.
• Check the Ethernet cables between the Bravo Platform and computer (and to and from hub if applicable).If the problem persists, contact Agilent Technical Support. Problems with the Bravo servo motors If the problem persists, contact Agilent Technical Support.