|
Depending on the type of device, the teachpoint setting procedure can vary. This topic provides basic teachpoint setting concepts: how to use the supplied teaching jig or the desired labware to set, verify, and edit Direct Drive Robot teachpoints.
You can have multiple teachpoint files for each Direct Drive Robot if a device needs to serve multiple purposes. For example, a platepad can be taught as a platepad or as a regripping station. You can also have multiple teachpoint files for the system if a device is only used in some protocols but not others. For example, if a device is installed on a docking table and can be removed when it is not in use.
The teachpoint file is referenced by a profile. For information about profiles, see Creating Direct Drive Robot profiles.The workflow for setting a Direct Drive Robot teachpoint is as follows:
Optional. Designate a teachpoint as a regrip station. Optional. Selecting the All regrip stations are safe from above option.
• You have the provided Direct Drive Robot teaching jig. If size restrictions at teachpoints prevent the use of the teaching jig, use the labware intended for the location. See Setting teachpoints using a labware.
IMPORTANT If size restrictions at a teachpoint prevent the use of the teaching jig, use the labware intended for the location. For information, see Setting teachpoints using a labware.
1 In DDR Diagnostics, click the Jog/Move tab, and then click Move to safe zone. The robot moves into its safe zone.
2
3
4 Click Teach Mode. The blue light on the robot hand turns on to indicate that it is in the teach mode. You should be able to move the robot arm without resistance.
If it is not possible to access the blue button on the robot hand, you can click the New teachpoint button in the Teachpoints tab.
4 To view the entire teachpoints table, click the(Maximize/Minimize) button. The table expands so that all of the columns are displayed in the tab.
Double-click the Name box and type a new name for the teachpoint. Use a name that describes the location, such as the device name (for example, PlateLoc - 1).
•
•
a Right-click the teachpoint in the teachpoints table, and then click Show robot position.
c Click OK to exit the dialog box and, if necessary, set or correct the A1-well orientation setting.
IMPORTANT The approach height value works in conjunction with the approach distance to clear obstacles near or at the teachpoint. Therefore, before setting the approach height value, you should determine the approach distance value to use. See Setting the approach distance.Agilent Technologies recommends the following Approach Ht (with labware) values:
Vertical Pipettor shelf with tipbox * Required approach height for the deviceDouble-click in the Approach Ht box and type the desired value. The default value for both parameters is 15 mm.
For example, you can use the approach distance parameter so that the robot grippers can enter and exit a device such as the Microplate Centrifuge without bumping into the sides of the narrow entryway.
Double-click in the Approach Distance box and type the desired value. The default value is 75 mm for the landscape orientation and 125 mm for the portrait orientation.The robot gripper offset range parameters tell the Direct Drive Robot where to grip a given labware type when transferring it from one device to another. Each device within a system can pose different accessibility challenges when transferring the labware. For example, some devices can have a deep, recessed area, whereas others are more flat. Some devices might have tall flanges that make it impossible for the robot to grip a microplate at a very low point.The Direct Drive Robot uses three types of gripper offset ranges to determine the most compatible grip position for the labware type:
1 Gripper offset range for the labware. Specified on the Direct Drive Robot tab in the Labware Editor for each labware definition.
2 Gripper offset range for pick-location device. Specified in the DDR Diagnostics software for a given device (device A).
3 Gripper offset range for place-location device. Specified in the DDR Diagnostics software for a given device (device B).Note: To catch and correct potential labware transfer errors, Agilent Technologies recommends that you verify the teachpoints and perform a dry run.
• Minimum gripper offset. The vertical distance (mm) from the teachpoint to the lowest point where the robot grippers can hold the microplate securely. The default value is 0 mm.
• Maximum gripper offset. The vertical distance (mm) from the teachpoint to the highest point where the robot grippers can hold the microplate securely. The default value is 10 mm.
IMPORTANT For regrip stations, make sure the gripper offset range can accommodate all types of labware.
• Position Z. The z-axis coordinate of the robot, measured in millimeters.
• Waist. The waist joint coordinate.
• Elbow. The elbow joint coordinate.
• Wrist. The wrist joint coordinate.Note: You can also use the jog method when fine-tuning. For instructions, see Editing existing teachpoints.
• Delidding at a Lid Hotel Station or a Vacuum Delid Station
• Relidding at a Lid Hotel Station
1 Select the appropriate action from the Pick Custom Action list.
• Note: You can select a custom action for each storage bay in the Lid Hotel Station. Do not assign a custom action for the platepad at the top of the station.
•
2 Select the appropriate action from the Place Custom Action list.
• Delid labware at the Lid Hotel Station or the Vacuum Delid Station. From the Place Custom Action list, select either Lid Hotel delid or Vacuum delid.Note: You can select a custom action for each storage bay in the Lid Hotel Station. Do not assign a custom action for the platepad at the top of the station.
When you create or modify teachpoints without saving them, the changes are kept in computer memory only. Saving the teachpoints in DDR Diagnostics updates the teachpoints file and the information in the firmware.Note: Saving teachpoints adds new teachpoints or updates only those teachpoints that have been modified in DDR Diagnostics. Saving teachpoints does not overwrite all of the existing teachpoints. For example, a teachpoint file contains teachpoints A and B. You modify teachpoint B and add teachpoint C. When you save the teachpoints, teachpoint C is added, and teachpoint B is updated. Teachpoint A is not touched.
1 Click Save teachpoints.
2 Click OK to close the dialog box and return to the teachpoints table. Notice that error and warning symbols appear in the teachpoints table.
4 Fix the errors, and then click Save teachpoints.Be sure to remove the teaching jig from the robot grippers when you are finished. See Installing and removing the teaching jig in the robot grippers.The All regrip stations are safe from above option enables the robot to rotate its wrist above the regrip station while it prepares to change labware orientation.After you have finished setting all the teachpoints, review the list in the teachpoints table. You can select the All regrip stations are safe from above option in the Profiles tab if:
• There is sufficient clearance above all regrip stations to permit the robot to rotate its wrist during a labware orientation regrip process.
Robot Diagnostics Direct Drive Robot axes