Vertical Pipetting Station User Guide : Using Diagnostics software : Using the Approach and Move commands

Using the Approach and Move commands
About this topic
This topic describes the Approach and Move commands. Typically, you use these commands when editing teachpoints to move the head quickly to specific locations on the Vertical Pipetting Station shelves.
 
IMPORTANT Only administrators and experienced personnel should use these procedures.
IMPORTANT To perform an immediate stop of the pipette head, click Stop motors.
Approaching a shelf
The Approach command moves the pipette head to a specified stopping distance above the teachpoint for the selected shelf.
To approach a shelf:
1
2
On the Jog/Teach tab, select the shelf that you want to approach.
3
In the Approach height box, type a safe stopping distance (mm) above the teachpoint.
4
Click Approach. The pipette head moves to the specified location above the saved teachpoint.
Moving to a teachpoint
The Move command moves the pipette head to the teachpoint for the selected shelf.
To move to a shelf teachpoint:
1
2
On the Jog/Teach tab, select the shelf.
3
Click Move. The pipette head moves to the saved teachpoint for the shelf.