BenchCel User Guide : Quick reference : BenchCel Diagnostics - General Settings tab
BenchCel Diagnostics - General Settings tab
You use the General Settings tab to adjust various settings. See Changing the general settings.
 
 
Parameters
Description
Stack Settings area
Rack sensor threshold
 
Any value at or higher than this means a rack is present. Below this value means there is no rack.
Low pressure threshold (psi)
The value at or below which the system displays a low-pressure warning.
Default value: 30 psi
Note: If this value is too low, the device could drop microplates without generating a low air-pressure warning.
Home robot after protocol run finishes
The option to return the robot to its home position after the protocol run is finished.
This option applies to all BenchCel devices that are controlled by the same computer running the VWorks software.
Speed settings
The command that opens the Speed Settings dialog box, which enables you to set each speed (Slow, Medium, and Fast) as a percentage of the factory-set maximum speed.
Show current settings
The command that displays the parameter values that were last saved.
Write settings to file
The command that allows you to save the Advanced Settings information to a file.
Apply
The command that applies the changes so that the new values are used.