![]() |
![]() |
![]() |
Step | Description | ||
---|---|---|---|
1 | Set dilution parameters | Enter values for the following inputs: | Default value (Range) |
Final concentration of dilution 1 | 0 (>0) | ||
Optional volume (µL) to be added | 0 (0 to < Final volume of dilution series) | ||
Final volume (µL) of dilution series | 0 (5–500) | ||
Direction of dilution | top to bottom (top to bottom, bottom to top, left to right, right to left) | ||
Number of replicates | 1 (1–5) | ||
Mix cycles after each dilution | 10 (0–100) | ||
Change pipette tips after each dilution | Yes (No/Yes) | ||
Pre-wet pipette tips before transfer | Yes (No/Yes) | ||
Blowout volume (µL) | 5 (0–50) | ||
Liquid class | Automatic (preset or custom) | ||
Serial dilution plate | |||
Diluent reservoir | |||
2 | Define dilution concentrations | Enter the final target concentration for each serial dilution step. | |
3 | Select the starting well of the dilution series | Click Select Starting Well, and then click the cell in the Serial Dilution Plate map where the replicate 1 of dilution 1 will be located. The setup tool will display the plate layout based on the input provided in steps 1 and 2. | |
4 | Calculate the sample and diluent volumes | Click Calculate Volumes. The setup tool calculates the required volumes of sample and diluent based on the input provided in steps 1 to 3. | |
5 | Correct these errors | Resolve any errors that the setup tool highlights. | |
6 | Create the method file | Click Create Serial Dilution Method. The setup tool creates a comma-separated value (.csv) text file that you use to run the utility. After you save the method, the Method Setup Tool closes. When you select the method in the Serial Dilution utility, the form displays the corresponding labware selections and diluent preparation instructions. Follow the instructions that appear in the form to prepare the serial dilution and diluent plates. |
Input | Description |
---|---|
Final concentration of dilution 1 | |
Concentration units | In the Input column, verify that the unit of measure for the sample concentration is correct. If necessary, type the correct units (default is mg/mL). This input is used to populate other sections of the form. This value has no impact on the subsequent data or volume calculations. |
Concentration value | In the Value column, type the final concentration of the most concentrated level in the serial dilution series (e.g., Dilution 1 in Step 2). |
Optional volume (µL) to be added | The volume to be added manually or using the Reagent Transfer utility after the Serial Dilution run, for example, an internal standard. The set up tool uses this value to reduce the volume of the serial dilution to the final volume minus the optional volume to be added. The set up tool also uses this value to increase the concentration of each serial dilution step (see Post Run cells in Step 2) so that when the optional volume is added, the final serial dilution volumes and concentrations match the desired final serial dilution series volumes and concentrations. ![]() For example, if the Final volume = 50 µL and Optional volume to add later = 10 µL, the Serial Dilution steps will have a volume of 40 µL after the run, at which point the optional 10 µL will be added for a total volume of 50 µL. Default: 0 (µL) Range: 0–50 (µL) |
Final volume (µL) of dilution series | The final volume of each step in the dilution series. This includes the Optional volume (µL) to be added. Note: The practical range is approximately 20–500 µL, depending on the number and concentration of the dilution steps. The high end of the range is limited by well capacities and the high volume transfers required during the early steps of the dilution series compared to the final volume of the dilution steps. The low end of the range is limited by the requirement that the transfers be 5 µL or greater to ensure high precision and accuracy. Large-volume dilution series require more mix cycles (up to 100 for the largest volume serial dilutions) because of the limited volume that can be used for each mix cycle. The results for large volume dilution series may not be as accurate as smaller volume dilution series. Default: 0 (µL) Range: 5–500 (µL) |
Direction of dilution | The direction in which the dilution series should progress. The dilution series follows a serpentine pattern. See the example in Step 3) Select the starting well of the dilution series. Default: top to bottom Options: top to bottom, bottom to top, left to right, right to left |
Number of replicates | The number of replicates of each dilution step in the dilution series. Note: This cell turns yellow if you select 5 replicates. The yellow cautions you that the serial dilution may not fit on the serial dilution plate, depending on the number of steps in the dilution series. The maximum number of dilution steps with 5 replicates is as follows: • 16 for top-to-bottom or bottom-to-top dilution series • 12 for left-to-right or right-to-left dilution series The Number of replicates value determines the total number of wells that will be prepared for each step in the dilution series. For example, if the number of replicates is set to 2, then on run completion, there will be two wells containing each concentration in the series. Default: 1 Range: 1–5 |
Mix cycles after each dilution | The number of aspirate-and-dispense cycles used to mix the contents of the wells in the serial dilution plate. Default: 10 Range: 0–100 |
Change pipette tips after each dilution | The option to change pipette tips before transferring the sample for each dilution step. Default: Yes Options: Yes, No |
Pre-wet pipette tips before transfer | The option to wet the pipette tips with the sample before aspirating the sample. Note: Prewetting the pipette tips is a common pipetting technique that can increase accuracy in certain situations. Default: Yes Options: Yes, No |
Blowout volume (µL) | The volume of air to be drawn into the pipette tips before aspirating the sample. After dispensing the transferred sample, the volume of air remaining in the pipette tips is dispensed (blown out) while the tips are still in the wells. The blowout is followed by a tip touch on the east and west sides of the well. If the Blowout volume is set to 0, no blowout or tip touches will occur. Default: 5 (µL) Range: 0–50 (µL) |
Liquid class | The pipetting parameters (for example, aspiration and dispense speed). The liquid class selection strongly influences the pipetting precision and accuracy. This choice only controls the aspiration and dispense of the samples and diluent. The mixing speed of the dilutions is fixed at 300 µL/sec. Options: • <Automatic> (default). Automatically assigns one of the following liquid classes, based on the volume being transferred. 0–20 µL (AM_250uLTipsLowVol) > 20 µL (AM_250uLTipsHighVol) These are good general-purpose liquid classes for most reagents that are used with the AssayMAP system. • Slow Flow (5 µL/sec). A slower flow rate is better for viscous solutions. • Fast Flow (100 µL/sec). A faster flow rate may improve performance for high organic solutions. If these flow rates do not provide the desired performance, you may enter a custom liquid class. To create a custom liquid class, you use the VWorks Liquid Library Editor. For details on how to open the VWorks Liquid Library Editor, see To create a custom liquid class:. To enter a custom liquid class: In the Liquid Class box, type the liquid class name exactly as it appears in the VWorks Liquid Library Editor. Note: The estimated run time (Status box area) may not be accurate for methods that use a custom liquid class. |
Serial dilution plate | The labware that you are placing on the plate riser at deck location 6. For options, see Labware. For the capacity limits and the volume overage recommended for each labware, see the Labware Reference Guide in the Literature Library page of the workbench. Note: The intermediate volumes required to generate the serial dilution will be larger than the final volume of the serial dilution. If you have chosen a plate type whose capacity is lower than required, you will be alerted in steps 4 and 5. |
Diluent reservoir | The labware that you are placing on the plate riser at deck location 2. For options, see Labware. For the capacity limits and the volume overage recommended for each labware, see the Labware Reference Guide in the Literature Library page of the workbench. Be sure to choose a diluent reservoir that can hold the volume required for the serial dilution. The volume of diluent and the location where it should be placed will be given in step 7 of the Method Setup Tool. The plate type chosen must have enough well capacity to hold the dilution indicated in step 7 plus the required well overage. For volume guidelines, see Preparing the diluent and samples. Note: If the selected labware cannot hold the volume of diluent required for the serial dilution, an error message will appear in steps 4 and 5. |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |