![]() |
Action | Deck location |
---|---|
a Uses the bare probes to aspirate 10 µL from the Reagent plate into the syringes. | 5 |
b Dispenses the 10 µL into the Sample plate. | 4 |
c Performs three mix cycles in the Sample plate to mix the contents. | 4 |
d Performs three syringe washes at the wash station. | 1 |
e Continues on to Addition Steps 2 to 5. | varies |
![]() |
Action | Deck location |
---|---|
a Sets the Peltier Thermal Station to 37 °C. Note: The heater is activated before the liquid transfer because it takes about 5 minutes for the temperature to reach the set point. | 4 |
b Uses the bare probes to aspirate 10 µL from the Reagent plate into the syringes. | 5 |
c Dispenses the 10 µL into the Sample plate. | 4 |
d Performs three mix cycles in the Sample plate to mix the contents. | 4 |
e Starts the incubation timer when the Peltier Thermal Station reaches 37 °C. | – |
f Performs three syringe washes at the wash station. | 1 |
g Remains idle for the remainder of the 45-minute incubation time. | – |
h Continues on to Addition Steps 2 to 5. | varies |
![]() |
Action | Deck location |
---|---|
a Uses the bare probes to aspirate 10 µL from the Reagent plate into the syringes. | 5 |
b Dispenses the 10 µL into the Sample plate. | 4 |
c Performs three mix cycles in the Sample plate to mix the contents. | 4 |
d Performs three syringe washes at the wash station. | 1 |
e A Pause message appears on the screen and the AssayMAP Bravo Platform stops executing new tasks. The action remains paused until you click the Continue button that appears on the screen. | – |
f Continues on to Addition Steps 2 to 5. | varies |
![]() |
Action | Deck location |
---|---|
a Sets the Peltier Thermal Station to 60 °C. Note: The heater is activated before the liquid transfer because it takes about 5 minutes for the temperature to reach the set point. | 4 |
b Uses the bare probes to aspirate 10 µL from the Reagent plate into the syringes. | 5 |
c Dispenses the 10 µL into the Sample plate. | 4 |
d Performs three mix cycles in the Sample plate to mix the contents. | 4 |
e Starts the incubation timer when the Peltier Thermal Station reaches 60 °C. | – |
f Performs three syringe washes at the wash station. | 1 |
g Remains idle for the remainder of the 60-minute incubation time. | – |
h A Pause message appears on the screen and the AssayMAP Bravo Platform stops executing new tasks. The action remains paused until you click the Continue button that appears on the screen. Note: The Peltier Thermal Station remains at temperature (60 °C in this case) until the Pause message is cleared. | – |
i Continues on to Addition Steps 2 to 5. | varies |
![]() |
Action | Deck location |
---|---|
a Sets the Peltier Thermal Station to 60 °C. Note: The heater is activated before the liquid transfer because it takes about 5 minutes for the temperature to reach the set point. | 4 |
b Uses the bare probes to aspirate 10 µL from the Reagent plate into the syringes. | 5 |
c Dispenses the 10 µL into the Sample plate. | 4 |
d Performs three mix cycles in the Sample plate to mix the contents. | 4 |
e Performs three syringe washes at the wash station. | 1 |
f A Pause message appears on the screen and the AssayMAP Bravo Platform stops executing new tasks. The action remains paused until you click the Continue button that appears on the screen. Note: The Peltier Thermal Station remains at temperature (60 °C in this case) until the Pause message is cleared. | – |
g Continues on to Addition Steps 2 to 5. | varies |
Scenario | Uses Peltier Thermal Station? | Start temperature (°C) | Target temperature (°C) | Ramp time (min), approximate |
---|---|---|---|---|
1 | Yes | 25 | 60 | < 5 |
2 | Yes | 25 | 10 | < 5 |
3 | Yes | 60 | 10 | < 10 |
4 | Yes | 37 | 25 | < 5 |
5 | No | 37 | 25 | > 15 |
6 | No | 60 | 25 | > 30 |
7 | No | 10 | 25 | > 20 |
![]() |
![]() |
Error check | Description |
---|---|
Sample Plate Capacity | Compares the Initial Sample Volume input to the Sample Plate Labware selection to verify that it does not exceed the labware volume capacity. |
Reagent Plate Capacity | Compares each Reagent Addition Volume input to its respective Labware Selection to ensure that it does not exceed the labware volume capacity. |
Cumulative Volume | Calculates the cumulative digest volume (Initial Sample Volume + sum of Reagent Addition Volumes) to ensure that the final digest volume does not exceed the Sample Plate Labware volume capacity. |
Sample Plate Pipette Tip Compatibility | If a deep-well plate type is selected for the Sample Plate Labware, each addition step is verified to ensure that the Use Pipette Tips for Addition? option is selected. |
Reagent Plate Pipette Tip Compatibility | If a deep-well plate type is selected for any of the five reagent Labware Selections, the corresponding Use Pipette Tips for Addition? setting is verified to ensure that pipettes will be used for the transfer. |
No Addition Steps | Checks to see that at least one reagent addition step has been specified by ensuring that not all of the Reagent Names settings have been set to NONE. |
Method Status error message | Reason for error | How to correct the error |
---|---|---|
File chosen is not a valid Method file. | The method file selected is for a different application. | Click Select Method, and then select a valid method file. |
File type is incorrect. Please choose a correct Method file. | The file selected is not a method file. | Click Select Method, and then select a valid method file. |
Error: Total volume will exceed Sample Labware capacity (0 µL). Please select a larger capacity Sample Labware. | No labware is selected for the Sample Plate. | Under Input Sample Settings, select the labware that you are using from the Sample Plate Labware list. |
Error: Total volume will exceed Sample Labware capacity (n µL). Please select a larger capacity Sample Labware. | The initial sample volume exceeds the capacity of the selected labware type. | Do one of the following: • Select a labware type with a larger capacity from the Sample Plate Labware list. • Reduce the value in the Starting Sample Volume box to a number that is within the capacity of the current labware selection. For a list of maximum volumes by labware type, see Labware. |
Error: Sample Plate is a type of deep well labware. Tips are needed in all steps. | Some of the addition steps specify the use of the AssayMAP bare probes, which cannot access the bottom of most deep-well plates. Pipette tips are required for addition steps that include a deep-well plate on the AssayMAP Bravo Platform. | Do one of the following: • Select the Use Tips for Addition check box for each Addition step to ensure that pipette tips are used for the transfers instead of the AssayMAP probes. • If you do not want to use pipette tips for the transfers, ensure that no deep-well plates are selected in the Sample Plate Labware list and the Labware Selection lists for the Addition steps. |
Error: Check Step 1 for volume/labware/tip combination. | The Volume input for one or more of the Addition Steps is greater than the volume capacity for the selected labware type. | Do one of the following for the corresponding Addition step: • Select a labware type with a larger capacity from the Labware Selection list. • Reduce the value in the Addition Volume box to a number that is within the capacity of the current labware selection. For a list of maximum volumes by labware type, see Labware. |
Error: Check Step 1 for volume/labware/tip combination. Error: Total volume will exceed Sample Labware capacity. Please select a larger capacity Sample Labware. | The cumulative volume in the sample plate, after all addition steps have completed, would be greater than the capacity of the selected Sample plate labware type. | Do one of the following: • Select a labware type with a larger capacity from the Sample Plate Labware list. • Reduce the values in the Starting Sample Volume box and/or the Addition Volume boxes to volumes that are within the capacity of the current labware selection. For a list of maximum volumes by labware type, see Labware. |
There are no reagent steps included on the form. Please double check and start the protocol again. | The Addition Name for all five addition steps is set to None, which causes all steps to be skipped. | For each addition step you want to include, select or type a name in the Addition Name box. Ensure that you have entered settings for at least one addition step in the Input Addition Step Settings area. |
Error: Incubation temperature for Addition step 1 is out of range. | The Incubation Temperature (°C) is set to a value that is outside of the allowable range (4 to 100 °C). | If you want to specify an on-deck incubation, enter a value within the allowable range (4 to 100 °C) in the Incubation Temperature (°C) box. If you do not want to specify a temperature, select OFF. ![]() |
Error: No labware associated with Addition step 1. | The Labware Selection is set to No Labware for one or more addition steps. | For each addition step you want to include, select the corresponding labware from the Labware Selection list. |
Error Check Step 1 for volume/labware/tip combination. | A deep-well plate is selected for one or more of the addition steps, but these steps are specified to use AssayMAP Probes, which cannot access the bottom of most deep-well plates. | Do one of the following: • Select the Use Tips for Addition check box for each Addition step to ensure that pipette tips are used for the transfers instead of the AssayMAP probes. • If you do not want to use pipette tips for the transfers, ensure that no deep-well plates are selected in the Labware Selection lists for the Addition steps. |
Error: Addition step 1 labware is a reservoir, lids cannot be used. | The selected labware type cannot have a lid, but the Plate Lidded setting is set to Lid. | Do one of the following: • Select No Lid in the Plate Lidded box for this Addition step. • Select a type of labware that can use a lid in the Labware Selection list for this Addition. |
![]() |
Protocol step | Head moves to deck location... | Action |
---|---|---|
Starting protocol | 2 | Parks all cartridges that might have been loaded on the head from a previously aborted protocol. |
1 | Dispenses any liquid remaining in the syringes into the wash station. | |
Addition step 1 | 1 | Washes the syringes. |
1 | Sets the Peltier Thermal Station to 60 °C and pauses until the temperature reaches 55 °C. | |
5 | Aspirates air into the syringes while moving to location 5. | |
5 | Prewets the syringes by cycling up and down 3 times. Goes to the top of the wells and dispenses air and then performs a tip touch. Aspirates 20 µL of Denaturation Mixture. | |
4 | Picks up the lid. | |
3 | Drops off the lid. | |
4 | Dispenses 20 µL of Denaturation Mixture, and then cycles up and down 15 times to mix the solution. Goes to the top of the wells and dispenses air and then does a tip touch. | |
3 | Picks up the lid. | |
4 | Drops off the lid. | |
1 | Dispenses any air remaining in the syringes, and then washes the syringes. Note: The protocol pauses until you click Continue in the Scheduler Paused window. | |
Addition step 2 | 1 | Sets the Peltier Thermal Station to 25 °C and pauses until the temperature reaches 30 °C. |
6 | Picks up the lid. | |
3 | Drops off the lid. | |
6 | Aspirates air into the syringes while moving to location 6. | |
6 | Prewets the syringes by cycling up and down 3 times. Goes to the top of the wells and dispenses air and then performs a tip touch. Aspirates 5 µL of Alkylant. | |
3 | Picks up the lid. | |
6 | Drops off the lid. | |
4 | Picks up the lid. | |
3 | Drops off the lid. | |
4 | Dispenses 5 µL of Denaturation Mixture, and then cycles up and down 15 times to mix the solution. Goes to the top of the wells and dispenses air and then does a tip touch. | |
3 | Picks up the lid. | |
4 | Drops off the lid. | |
1 | Dispenses any air remaining in the syringes, and then washes the syringes. | |
1 | Waits for 30-minute incubation. | |
Addition step 3 | 1 | Sets the Peltier Thermal Station to 25 °C and pauses until the temperature reaches 20–25 °C. |
7 | Aspirates air into the syringes while moving to location 7. | |
7 | Prewets the syringes by cycling up and down 3 times. Goes to the top of the wells and dispenses air and then performs a tip touch. Aspirates 140 µL of Diluent Mixture. | |
4 | Picks up the lid. | |
3 | Drops off the lid. | |
4 | Dispenses 140 µL of Diluent Mixture, and then cycles up and down 15 times to mix the solution. Goes to the top of the wells and dispenses air and then does a tip touch. | |
3 | Picks up the lid. | |
4 | Drops off the lid. | |
1 | Dispenses any air remaining in the syringes, and then washes the syringes. | |
1 | Waits for 0-minute incubation. | |
Addition step 4 | 1 | Sets the Peltier Thermal Station to 37 °C. |
1 | Moves the syringes up and down for 4 cycles with a wicking across the chimneys after every cycle. | |
2 | Mounts the pipette tips on the head. | |
8 | Aspirates air into the syringes while moving to location 8. | |
8 | Prewets the pipette tips by cycling up and down 3 times. Goes to the top of the wells and dispenses air and then performs a tip touch. Aspirates 5 µL of Protease. | |
4 | Picks up the lid. | |
3 | Drops off the lid. | |
4 | Dispenses 5 µL of Protease, and then cycles up and down 15 times to mix the solution. Goes to the top of the wells and dispenses air and then does a tip touch. | |
3 | Picks up the lid. | |
4 | Drops off the lid. | |
1 | Dispenses any air remaining in the syringes, and then washes the pipette tips. | |
2 | Parks the pipette tips in the seating station. | |
1 | Waits for 180-minute incubation. | |
Addition step 5 | 1 | Sets the Peltier Thermal Station to 25 °C. |
9 | Aspirates air into the syringes while moving to location 9. | |
9 | Prewets the syringes by cycling up and down 3 times. Goes to the top of the wells and dispenses air and then performs a tip touch. Aspirates 20 µL of Acidification Reagent. | |
4 | Picks up the lid. | |
3 | Drops off the lid. | |
4 | Dispenses 20 µL of Acidification Reagent, and then cycles up and down 15 times to mix the solution. Goes to the top of the wells and dispenses air and then does a tip touch. | |
3 | Picks up the lid. | |
4 | Drops off the lid. | |
1 | Dispenses any air remaining in the syringes, and then washes the syringes. | |
1 | Waits for 0-minute incubation. |