Control or indicator | Description |
---|---|
Location (Collection) | Specifies the Bravo deck location to be used for the selected process command. If the selected Command to execute is Assemble Vacuum, this list is labeled Collection, and the selection controls the location for the Collection plate. |
Labware at selected location (Collection) | Specifies the labware to be used for the selected process command. Note: To add a labware selection to the list, click Open labware editor, and use the Labware Editor. |
Location 2 (Filter) | Specifies the Bravo deck location to be used for the selected process command. If the selected Command to execute is Assemble Vacuum, the list is labeled Filter and the selection controls the location for the Filter plate. |
Labware at selected location (Filter) | Specifies the labware to be used at location 2 for the selected process command. Note: To add a labware selection to the list, click Open labware editor, and use the Labware Editor. |
Graphical display of Bravo deck | An interactive display that provides the following: • Alternative way to specify the target location. • Visual display of the type of platepad and labware configured for each location. |
Control or indicator | Description |
---|---|
Open labware editor | Opens the Labware Editor. |
Open pipette technique editor | Opens the Pipette Technique Editor. |
Open liquid library | Opens the Liquid Library Editor. |
Control or indicator | Description |
---|---|
Plate graphical display | Provides an interactive display that changes based on your labware selection. For example, if you use a 96-tip head and a 96-well microplate, all the wells are selected (green) by default. If you use a 96-tip head and a 384-well microplate, only one quadrant of the wells are selected by default. To select a quadrant: Click a well in that quadrant. All the wells in that quadrant turn green, indicating the selected wells. Alternatively, right-click the graphical display to access the following shortcut menu commands: • Select all wells • Clear all selected wells • Select all wells in highlighted row • Clear all selected wells in highlighted row • Select all wells in highlighted column • Clear all selected wells in highlighted column |
Set head mode | Opens the Head Mode Selector dialog box. Series III disposable-tip heads and Bravo 96AM Head only. To use one row or column of channels instead of all channels in the head, use the controls in the dialog box to select which channels on the head to use. |
Parameter | Description |
---|---|
Command to execute | Provides the following list of tasks that you can run: • Tips On and Tips Off • Vacuum Filtration Station task parameters (Assemble Vacuum, Disassemble Vacuum, and Move and Filter Plate) Note: Other commands may be available depending on the installed accessories. For a complete description of task parameters, see the liquid-handling tasks in the VWorks Automation Control User Guide. |
Execute command | Performs the selected task command. Note: To stop a task before it finishes running, click Stop motors. To resume movement of the head, you can use the Enable all motors button on the Jog/Teach tab. |