Microplate Labeler User Guide : Installing the device : Connecting the air tubing: non-metric
Connecting the air tubing: non-metric
About this topic
For non-metric-based installations, such as those in the U.S., this topic explains how to connect the air tubing to the Microplate Labeler.
Before you begin
 
 
 
*Make sure that the air tubing coming into your Microplate Labeler is properly filtered for moisture. Moist air might shorten the life of your device. If the quality of your air supply is in question, contact Agilent Automation Solutions Technical Support.
 
Locate the air connection kit (non-metric), including tubing and fittings, provided with your device.
Connection procedure
To set up a direct connection, use the following procedure. If you have another device connected to the air supply, use the supplied quick-release T‑connector to connect to the existing air tubing.
Figure Connection panel: orange air-supply port (left) and pressure gauge (right)
 
To install the air tubing:
1 Cut the supplied 1/4-inch tubing to the required length.
2 If your air supply uses a threaded connection at the supply end, attach the 1/4-inch NPT fitting to the connection.
3 Plug one end of the tubing into your lab’s air supply.
4 Connect the tubing from the air source to the long end of the toggle switch assembly. Connect the short end of the toggle switch assembly to the tubing that connects to the Microplate Labeler.
5 Push the other end of the 1/4-inch tubing into the orange quick-release air-supply port at the back of the Microplate Labeler.
6 Test the installation by gently tugging the air tubing.
If you feel resistance, the line has been installed properly.
Related topics
 
For information about...
See...
How to install the device
Site requirements for installation
How to connect air tubing for metric-based installations
Setting the air pressure
Disconnecting air supply